Advanced Diagnostics Configuration
March 2016
Advanced Diagnostics Configuration
6.10.2
Test scope
The manually initiated
SMART Meter Verification
test can be used to verify the entire
flowmeter installation or individual parts such as the transmitter or sensor. This parameter
is set at the time that the
SMART Meter Verification
test is manually initiated. There are three
test scopes from which to choose.
All
Run the
SMART Meter Verification
test and verify the entire flowmeter installation. This
parameter results in the diagnostic performing the transmitter calibration verification,
sensor calibration verification, coil health check, and electrode health check. Transmitter
calibration and sensor calibration are verified to the percentage associated with the test
condition selected when the test was initiated. This setting applies to manually initiated
tests only.
Transmitter
Run the
SMART Meter Verification
test on the transmitter only. This results in the verification
test only checking the transmitter calibration to the limits of the test criteria selected when
the verification test was initiated. This setting applies to manually initiated tests only.
Sensor
Run the
SMART Meter Verification
test on the sensor only. This results in the verification test
checking the sensor calibration to the limits of the test criteria selected when the
SMART
Meter Verification
test was initiated, verifying the coil circuit health, and the electrode circuit
health. This setting applies to manually initiated tests only.
6.11
Continuous SMART Meter Verification
Continuous SMART Meter Verification can be used to monitor and verify the health of the
flowmeter system. The continuous SMART Meter Verification will not report results until 30
minutes after powering up to ensure the system is stable and to avoid false failures.
Continuous SMART Meter Verification is only available on the Field Mount Transmitter.
6.11.1
Test scope
Continuous SMART Meter Verification can be configured to monitor the sensor coils,
electrodes, transmitter calibration, and analog output. All of these parameters can be
individually enabled or disabled. These parameters apply to continuous SMART Meter
Verification only.
LOI menu path
Field Mount: Diagnostics, Diag Controls, Cont Meter Ver
Wall Mount: N/A
Traditional Fast Keys
1, 2, 1, 3
Device dashboard
2, 2, 6, 4
Содержание Rosemount 8750W
Страница 2: ......
Страница 96: ...84 Reference Manual 00809 0300 4750 Rev CA Operation March 2016 Operation ...
Страница 158: ...146 Reference Manual 00809 0300 4750 Rev CA Digital Signal Processing March 2016 Digital Signal Processing ...
Страница 168: ...156 Reference Manual 00809 0300 4750 Rev CA Maintenance March 2016 Maintenance ...
Страница 208: ...Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA 196 Product Certifications ...
Страница 209: ...197 Product Certifications Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA ...
Страница 212: ...Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA 200 Product Certifications ...
Страница 214: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 202 Wiring Diagrams ...
Страница 218: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 206 Wiring Diagrams ...
Страница 219: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 207 Wiring Diagrams ...
Страница 220: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 208 Wiring Diagrams ...
Страница 222: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 210 Wiring Diagrams ...
Страница 223: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 211 Wiring Diagrams ...
Страница 224: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 212 Wiring Diagrams ...
Страница 225: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 213 Wiring Diagrams ...
Страница 227: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 215 Wiring Diagrams ...
Страница 228: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 216 Wiring Diagrams ...
Страница 229: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 217 Wiring Diagrams ...
Страница 263: ......