164
Reference Manual
00809-0300-4750, Rev CA
March 2016
Table 9-3. Advanced Meter Verification Messages
Error message
Potential cause
Corrective action
8714i Failed
Transmitter calibration
verification test failed
•Verify pass/fail criteria
•Rerun SMART
™
Meter Verification (8714i) under no flow
conditions
•Verify calibration using 8714 Calibration Standard
•Perform digital trim
•Replace electronics board
Sensor calibration test failed
•Verify pass/fail criteria
•Rerun SMART Meter Verification (8714i)
•Perform sensor tests - see
Sensor coil circuit test failed
•Verify pass/fail criteria
•Rerun SMART Meter Verification (8714i)
•Perform sensor tests - see
Sensor electrode circuit test
failed
•Verify electrode resistance has a baseline (signature) value
from a full pipe baseline
•Verify test condition was selected properly
•Verify pass/fail criteria
•Rerun SMART Meter Verification (8714i)
•Perform sensor tests - see
4-20 mA loop
verification failed
(1)
Analog loop not powered
•Check 4-20 mA internal/external loop power switch - see
“Internal/external analog power” on page 38
•Check external supply voltage to the transmitter
•Check for parallel paths in the current loop
Transmitter failure
•Perform transmitter self test
•Perform manual analog loop test and D/A trim if necessary
•Replace the electronics board
Continuous Meter
Verification Error
(1)
Transmitter calibration
verification test failed
•Verify pass/fail criteria
•Run manual SMART Meter Verification (8714i) under no
flow conditions
•Verify calibration using 8714D Calibration Standard
•Perform digital trim
•Replace electronics stack
Sensor calibration test failed
•Run manual SMART Meter Verification (8714i)
•Perform sensor tests - see
Sensor coil circuit test failed
•Run manual SMART Meter Verification (8714i)
•Perform sensor tests - see
Sensor electrode circuit test
failed
•Run manual SMART Meter Verification (8714i)
•Perform sensor tests - see
•Verify electrode resistance has a signature value from a full
pipe baseline
Simulated Velocity Out
of Spec
(1)
Unstable flow rate during the
verification test or noisy process
•Run manual transmitter verification test with no flow and a
full pipe
Transmitter drift or faulty
electronics
•Verify transmitter electronics with 8714D Calibration
Standard. The dial on the 8714D should be set to 30 ft/s
(9.14 m/s). The transmitter should be set up with the
nominal calibration number (1000015010000000) and 5
Hz coil drive frequency.
•Perform an electronics trim using the 8714
•If the electronics trim doesn't correct the issue, replace the
electronics
Содержание Rosemount 8750W
Страница 2: ......
Страница 96: ...84 Reference Manual 00809 0300 4750 Rev CA Operation March 2016 Operation ...
Страница 158: ...146 Reference Manual 00809 0300 4750 Rev CA Digital Signal Processing March 2016 Digital Signal Processing ...
Страница 168: ...156 Reference Manual 00809 0300 4750 Rev CA Maintenance March 2016 Maintenance ...
Страница 208: ...Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA 196 Product Certifications ...
Страница 209: ...197 Product Certifications Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA ...
Страница 212: ...Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA 200 Product Certifications ...
Страница 214: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 202 Wiring Diagrams ...
Страница 218: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 206 Wiring Diagrams ...
Страница 219: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 207 Wiring Diagrams ...
Страница 220: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 208 Wiring Diagrams ...
Страница 222: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 210 Wiring Diagrams ...
Страница 223: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 211 Wiring Diagrams ...
Страница 224: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 212 Wiring Diagrams ...
Страница 225: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 213 Wiring Diagrams ...
Страница 227: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 215 Wiring Diagrams ...
Страница 228: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 216 Wiring Diagrams ...
Страница 229: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 217 Wiring Diagrams ...
Страница 263: ......