5.4
Rosemount 6888Xi Quick Start Wizard
When the Rosemount 6888Xi is first powered, a short wizard program guides you through
the basic setup. Once configured, the Rosemount 6888Xi retains the setup, and the wizard
will not repeat.
1. Apply power to the Rosemount 6888Xi.
Once boot-up is complete, the
Quick Start Wizard
screen appears. With a dual
channel Rosemount 6888Xi, the wizard runs for both channels in succession.
2. Press
Enter
to continue.
3. At the
Sensor Type
screen, use the
Up
and
Down
keys to select O
2
.
Do not select CO as this option is reserved for future use.
4. Press
Enter
to continue.
5. At the
Device Type
screen use the
Up
and
Down
keys to select HART
®
or FF
(FOUNDATION
™
Fieldbus), whichever applies.
6. At the
Auto Cal Device
screen, use the
Up
and
Down
keys to select the calibration
method to be used. The methods are defined as follows:
•
None - Manual calibration with the standard probe housing configuration
•
SPS - Automatic calibration with the standard probe housing configuration using
the Rosemount SPS 4001B
•
IMPS - Automatic calibration with the standard probe housing configuration using
the Rosemount IMPS
•
Integral - Automatic calibration with the integral autocal probe housing
configuration
7. Press
Enter
to continue.
NOTICE
If SPS, IMPS, or Integral is selected, you must still configure automatic calibration as
On. Other parameters, such as test gas values and gas times, should be verified as well.
Refer to the
Rosemount 6888Xi Reference Manual
for calibration setup details.
8. When prompted by
Setup Correct?
, use the
Up
and
Down
keys to select Yes.
If you select No, the wizard restarts.
9. Press
Enter
to continue.
The Rosemount 6888Xi displays several screens while saving the configuration, resets
itself, and then returns to the main screen.
5.5
Re-initiating Rosemount 6888Xi wizard
To re-initiate the wizard setup, you must reset the I/O board to default factory conditions.
1. Apply power to the Rosemount 6888Xi.
2. When the main screen appears, press
Menu
several times until the
System
menu
appears. Use the
Up
and
Down
keys to select Configure IOB. Press
Enter
to continue.
3. When the
Configure IOB
screen appears, use the
Up
and
Down
keys to select I/O
Board 1. With a dual channel Rosemount 6888Xi, you can reset and reconfigure either
or both I/O boards. Press
Enter
to continue.
Instruction Manual
Power up
00809-0100-4891
July 2018
Instruction Manual
49
Содержание Rosemount 6888C
Страница 10: ...General information Instruction Manual July 2018 00809 0100 4891 10 6888C ...
Страница 14: ...Typical system package Instruction Manual July 2018 00809 0100 4891 14 6888C ...
Страница 20: ...Specifications Instruction Manual July 2018 00809 0100 4891 20 6888C ...
Страница 46: ...Install Instruction Manual July 2018 00809 0100 4891 46 6888C ...
Страница 55: ...Instruction Manual Calibration 00809 0100 4891 July 2018 Instruction Manual 55 ...
Страница 56: ...Calibration Instruction Manual July 2018 00809 0100 4891 56 6888C ...
Страница 58: ...Figure 7 2 Rosemount 6888 Xi continued HART menu trees Instruction Manual July 2018 00809 0100 4891 58 6888C ...
Страница 60: ...Figure 7 4 Field Communicator HART menu trees Instruction Manual July 2018 00809 0100 4891 60 6888C ...
Страница 61: ...Instruction Manual HART menu trees 00809 0100 4891 July 2018 Instruction Manual 61 ...
Страница 62: ...HART menu trees Instruction Manual July 2018 00809 0100 4891 62 6888C ...
Страница 78: ...Maintenance and service Instruction Manual July 2018 00809 0100 4891 78 6888C ...
Страница 82: ...Service support Instruction Manual July 2018 00809 0100 4891 82 6888C ...
Страница 87: ...Instruction Manual 00809 0100 4891 July 2018 Instruction Manual 87 ...