5.2
Power up the Rosemount 6888 Transmitter
with single/dual channel or single channel and
flame safety interlock Rosemount 6888Xi
Complete the following steps to power up the Rosemount 6888 Transmitter with single/
dual channel or single channel and flame safety interlock Rosemount 6888Xi.
1. Appy AC line power to the transmitter.
2. Apply AC line power to the Rosemount 6888Xi. Run the Quick Start Wizard as
described in
Rosemount 6888Xi Quick Start Wizard
on page 49. At the
Auto Cal
Device
screen, select the calibration method based on the Rosemount 6888
Transmitter as follows:
a. Standard probe housing configuration - Select None, SPS, or IMPS as appropriate.
Do not select Integral or calibration will not be possible.
b. Integral autocal probe housing - Select Integral only. If Integral is not selected,
calibration will not be possible.
3. Verify communications between the transmitter and the Rosemount 6888Xi.
The Rosemount 6888Xi display is preconfigured to display O
2
and cell temperature for
single channel configurations and both O
2
readings for dual channel configurations.
The transmitter probe takes approximately 45 minutes to warm up to the 736 °C (1357 °F)
heater setpoint. The 4-20 mA signal remains at a default value of 3.5 mA, and the O
2
reading remains at 0% through this warm-up period. After warm up, the probe begins
reading oxygen, and the 4-20 mA output is based on the default range of 0-10% O
2
.
If there is an error condition at startup, an alarm message is displayed on the Rosemount
6888Xi.
5.3
Power up the Rosemount 6888 direct
replacement probe (no electronics inside) with
traditional architecture Rosemount 6888Xi
Complete the following steps to power up the Rosemount 6888 direct replacement probe
with the Rosemount 6888Xi Electronics.
1. Apply AC line power to the Rosemount 6888Xi.
2. Run the Quick Start Wizard as described in
Rosemount 6888Xi Quick Start Wizard
on
page 49.
3. At the
Auto Cal Device
screen, select None, SPS, or IMPS as appropriate.
Do not select Integral or calibration will not be possible.
The transmitter probe takes approximately 45 minutes to warm up to the 736 °C (1357 °F)
heater setpoint. The 4-20 mA signal remains at a default value of 3.5 mA, and the O
2
reading remains at 0% through this warm-up period. After warm up, the probe begins
reading oxygen, and the 4-20 mA output is based on the default range of 0-10% O
2
.
If there is an error condition at startup, an alarm message is displayed on the Rosemount
6888Xi.
Power up
Instruction Manual
July 2018
00809-0100-4891
48
6888C
Содержание Rosemount 6888C
Страница 10: ...General information Instruction Manual July 2018 00809 0100 4891 10 6888C ...
Страница 14: ...Typical system package Instruction Manual July 2018 00809 0100 4891 14 6888C ...
Страница 20: ...Specifications Instruction Manual July 2018 00809 0100 4891 20 6888C ...
Страница 46: ...Install Instruction Manual July 2018 00809 0100 4891 46 6888C ...
Страница 55: ...Instruction Manual Calibration 00809 0100 4891 July 2018 Instruction Manual 55 ...
Страница 56: ...Calibration Instruction Manual July 2018 00809 0100 4891 56 6888C ...
Страница 58: ...Figure 7 2 Rosemount 6888 Xi continued HART menu trees Instruction Manual July 2018 00809 0100 4891 58 6888C ...
Страница 60: ...Figure 7 4 Field Communicator HART menu trees Instruction Manual July 2018 00809 0100 4891 60 6888C ...
Страница 61: ...Instruction Manual HART menu trees 00809 0100 4891 July 2018 Instruction Manual 61 ...
Страница 62: ...HART menu trees Instruction Manual July 2018 00809 0100 4891 62 6888C ...
Страница 78: ...Maintenance and service Instruction Manual July 2018 00809 0100 4891 78 6888C ...
Страница 82: ...Service support Instruction Manual July 2018 00809 0100 4891 82 6888C ...
Страница 87: ...Instruction Manual 00809 0100 4891 July 2018 Instruction Manual 87 ...