![Emerson Rosemount 3490 Series Скачать руководство пользователя страница 99](http://html.mh-extra.com/html/emerson/rosemount-3490-series/rosemount-3490-series_reference-manual_2400788099.webp)
Table 4-11: Auxiliary Functions for Relay Mode - Special alarms
Relay mode
Relay ops
Pump
efficiency
Pumped
volume
Relay run time No activity
Rising level
Set point (PV)
Yes
Yes
N/A
Yes
Yes
N/A
Set Point (SV)
Yes
Yes
N/A
Yes
Yes
N/A
Set Point (TV)
Yes
Yes
N/A
Yes
Yes
N/A
Set Point (FV)
Yes
Yes
N/A
Yes
Yes
N/A
Assist
Yes
Yes
Yes
Yes
Yes
Yes
Stby Com-off
Yes
Yes
Yes
Yes
Yes
Yes
Stby Split-off
Yes
Yes
Yes
Yes
Yes
Yes
Digital Inputs
Yes
N/A
N/A
Yes
Yes
N/A
Sampler
N/A
N/A
N/A
N/A
N/A
N/A
Rate change
Yes
N/A
N/A
Yes
Yes
N/A
Off
N/A
N/A
N/A
N/A
N/A
N/A
Alarm
N/A
N/A
N/A
N/A
N/A
N/A
Hi/Lo Alarm
N/A
N/A
N/A
N/A
N/A
N/A
Totalizer
N/A
N/A
N/A
N/A
N/A
N/A
Fault
N/A
N/A
N/A
N/A
N/A
N/A
Custom
N/A
N/A
N/A
N/A
N/A
N/A
PV Limits
Yes
N/A
N/A
Yes
Yes
N/A
On
N/A
N/A
N/A
N/A
N/A
N/A
Duty assist relay with common off points
This function requires two or more Assist mode relays. Two or more of these relays can be
energized at the same time (assisting), and they all de-energize at a common off point.
An Assist relay uses its On Point and Off Point parameters in a similar way to a Set Point
mode relay. A typical application would be emptying a wet well/lift station.
The worked example here illustrates how the function works. To keep the example simple,
the auto-sequencing options are not considered.
Example wet well/lift station application (emptying operation due to rising level)
using assist mode relays and common off points
Consider an application with two relay outputs, RL1 and RL2, connected to individual
pumps in a wet well/lift station. The control unit PV value (D800) is a level measurement in
meters (m).
, both Pump 1 and Pump 2 are off because the liquid level is at a satisfactory
level, below 5 m.
Reference Manual
Getting started
00809-0100-4841
January 2021
Emerson.com/Rosemount
99
Содержание Rosemount 3490 Series
Страница 4: ...4 ...
Страница 8: ...Introduction Reference Manual January 2021 00809 0100 4841 8 Emerson com Rosemount ...
Страница 16: ...Control unit overview Reference Manual January 2021 00809 0100 4841 16 Emerson com Rosemount ...
Страница 36: ...Installation Reference Manual January 2021 00809 0100 4841 36 Emerson com Rosemount ...
Страница 134: ...Getting started Reference Manual January 2021 00809 0100 4841 134 Emerson com Rosemount ...
Страница 158: ...Specifications and reference data Reference Manual January 2021 00809 0100 4841 158 Emerson com Rosemount ...
Страница 184: ...Additional features Reference Manual January 2021 00809 0100 4841 184 Emerson com Rosemount ...
Страница 189: ...Reference Manual 00809 0100 4841 January 2021 Emerson com Rosemount 189 ...