4
Troubleshooting and maintenance
The topics in this section provide information on identifying and resolving issues and
maintaining the level transmitter.
4.1
Error indicators
This table describes error scenarios for the level transmitter and possible solutions.
Indicator
Possible cause
Recommended fix
Err
Fault in the electronics.
Replace the unit.
nPrb
Probe detached from the unit; possibly
incorrect setting of the probe length.
Check whether the probe is still
attached to the unit. Check the
parameter
LEnG
.
SEnS
Measurement disturbed by heavy foam
build-up or turbulence.
•
Install the unit in a still pipe or
bypass (
)
•
Set or increment
dFo
Measurement disturbed by separation
layers (e.g. oil layer on water).
Remove the oil layer by suction, stir the
medium, verify the composition.
Probe or process connection soiled.
Clean the probe and the process
connection.
Installation conditions not adhered to.
•
•
Repeat or perform a tank
adjustment. (
)
Probe length or sensitivity (setting to
the medium) incorrect.
Correct settings then perform a tank
adjustment (
).
SCx + LED 7
SCx + LED 8
Flashing: short circuit in switching
output OUT1 or OUT2.
Remove the short circuit.
SC + LED 7
+ LED 8
Flashing: short circuit in both switching
outputs
Remove the short circuit.
PArA
Faulty data set
Restore factory default settings (See
Reset all parameters to factory default
settings
4.2
Output response in different operating states
These tables provide information on the response of outputs for different operating
states.
Operating state
OUT1
OUT2
Initialization
OFF
OFF
Quick Start Guide
Troubleshooting and maintenance
00825-0100-4326
January 2019
Quick Start Guide
43
Содержание Rosemount 326L
Страница 1: ...Quick Start Guide 00825 0100 4326 Rev 01 January 2019 Rosemount 326L Level Transmitter ...
Страница 2: ...2 ...
Страница 4: ...Contents Quick Start Guide January 2019 00825 0100 4326 4 Quick Start Guide ...
Страница 12: ...Introduction Quick Start Guide January 2019 00825 0100 4326 12 Quick Start Guide ...
Страница 24: ...Installation Quick Start Guide January 2019 00825 0100 4326 24 Quick Start Guide ...
Страница 42: ...Operation Quick Start Guide January 2019 00825 0100 4326 42 Quick Start Guide ...
Страница 46: ...Troubleshooting and maintenance Quick Start Guide January 2019 00825 0100 4326 46 Quick Start Guide ...
Страница 48: ...Setting ranges Quick Start Guide January 2019 00825 0100 4326 48 Quick Start Guide ...
Страница 50: ...Product certifications Quick Start Guide January 2019 00825 0100 4326 50 Quick Start Guide ...