To correctly transfer the microwave pulse to the tank with optimum transmission power,
the level transmitter needs a sufficiently large metal launching surface or launching plate.
For installation in closed metal tanks or metal bypass pipes, the tank lid or upper pipe
section serves as a launching surface. Use a sufficiently large fixing plate, metal plate or
something similar when installing in open metal tanks, plastic tanks, or metal tanks with
plastic lids. For more information, see
3.5.2
Operation with a bypass or still pipe
Use a bypass or a still pipe in applications like heavy foam build-up.
For more information about heavy foam build-up and turbulence, see
.
For more information about the minimum internal pipe diameter, see
.
For general installation instructions, see
3.5.3
Function check
After turning on the level transmitter, it is in operating mode. It measures and evaluates
level and generates output signals according to the set parameters.
Check that the level transmitter is operating correctly before using it in a production
environment.
3.5.4
Operation indication
These indicators show how the level transmitter is operating.
Indicator
Definition
---- continuous
Initialization phase after power on
Initially the level transmitter is not operational. Configure basic settings
to enable the level transmitter (see
).
----
Level below the active zone
Numerical value + LED
1
Current level in millimeters
Numerical value + LED
2
Current level in inches
Numerical value + LED
3
Current level in percent of the scaled measuring range
LED 7 / LED 8
Switching status OUT2/OUT1
FULL + numerical value
alternately
Level has reached or exceeded the maximum measuring range (=
overflow warning).
SIM + XXX
Simulation active. XXX = state to be simulated (See
)
S.On
Simulation stopped (See
)
Quick Start Guide
Operation
00825-0100-4326
January 2019
Quick Start Guide
39
Содержание Rosemount 326L
Страница 1: ...Quick Start Guide 00825 0100 4326 Rev 01 January 2019 Rosemount 326L Level Transmitter ...
Страница 2: ...2 ...
Страница 4: ...Contents Quick Start Guide January 2019 00825 0100 4326 4 Quick Start Guide ...
Страница 12: ...Introduction Quick Start Guide January 2019 00825 0100 4326 12 Quick Start Guide ...
Страница 24: ...Installation Quick Start Guide January 2019 00825 0100 4326 24 Quick Start Guide ...
Страница 42: ...Operation Quick Start Guide January 2019 00825 0100 4326 42 Quick Start Guide ...
Страница 46: ...Troubleshooting and maintenance Quick Start Guide January 2019 00825 0100 4326 46 Quick Start Guide ...
Страница 48: ...Setting ranges Quick Start Guide January 2019 00825 0100 4326 48 Quick Start Guide ...
Страница 50: ...Product certifications Quick Start Guide January 2019 00825 0100 4326 50 Quick Start Guide ...