Figure E-1: Typical SGM installation
A. Calibration gas selector three-way valve
B. Process/Calibration gas selector three-way valve
C. Outlet two-way valve
D. Process gas inlet two-way valve
E. Chamber fill valve
F. Purge two-way valve
G. Pressure regulator with gauge
H. Calibration gas 1 inlet
I. Calibration gas 2 inlet
J. Process gas sampling point
K. Tubing work (not shown in diagram)
L. Coalescent filter
M. Coalescent filter drain two-way valve
N. Particle filter
O. Flow meter
P. Transmitter
Q. Reference chamber pressure indicator
R. Relief valve
S. Drain
Procedure
1. Choose
Device Tools
→
Calibration
→
Gas Calibration
.
2. Set the measurement type for this device.
Option
Description
Specific Gravity (SG)
Gas density will be measured as specific gravity, and specific gravity will be
used for calibration.
Molecular Weight (MW)
Gas density will be measured as molecular weight, and molecular weight
will be used for calibration.
Relative Density (RD)
Gas density will be measured as relative density, and relative density will
be used for calibration.
Three-point calibration
Configuration and Use Manual
March 2021
MMI-20020954
184
Micro Motion Gas Specific Gravity Meters (SGM)