Table 14-2: Status alerts, causes, and recommendations
(continued)
Alert number
Alert title
Possible causes
Recommended actions
A037
Sensor Check
Failed
Known Density Verification
failed.
•
Check the subtest results and perform the
recommended actions.
•
Retry the test.
•
Contact Micro Motion.
A038
Time Period
Signal Out of
Range
The time period signal is
outside the limits for the
sensor type.
•
Check your process conditions against the values
reported by the device.
A100
mA Output 1
Saturated
The calculated mA output
value is outside the
configured range.
•
Check the settings of
Upper Range Value
and
Lower Range Value
. See
•
Check process conditions. Actual conditions may
be outside the normal conditions for which the
output is configured.
•
Check for foreign material in the process gas or
fluid, coating, or other process problems.
•
Verify that the measurement units are configured
correctly for your application.
A101
mA Output 1
Fixed
The HART address is set to a
non-zero value, or the mA
output is configured to send
a constant value.
•
Check whether the output is in loop test mode. If it
is, unfix the output.
•
Exit mA output trim, if applicable.
•
Check the HART address. If the HART address is
non-zero, you may need to change the setting of
mA Output Action
(
Loop Current Mode
).
•
Check whether the output has been set to a
constant value via digital communication.
A102
Drive
Overrange
The drive power (current/
voltage) is at its maximum.
•
Check the drive gain and the pickoff voltage. See
.
•
Check for foreign material in the process gas or
fluid, coating, or other process problems.
•
Check for fluid separation by monitoring the
density value and comparing the results against
expected density values.
•
Ensure that the sensor orientation is appropriate
for your application. Settling from a two-phase or
three-phase fluid can cause this alert.
A104
Calibration in
Progress
A calibration procedure is in
process.
•
Allow the procedure to complete.
A106
Burst Mode
Enabled
HART burst mode is
enabled.
•
No action required.
•
If desired, you can set
Alert Severity Level
to
Ignore.
Configuration and Use Manual
Troubleshooting
MMI-20020954
March 2021
Configuration and Use Manual
131