3. Ensure that the matrix is configured correctly.
4. Adjust the concentration value by applying an offset. See
Adjust concentration measurement with
.
14.13 Milliamp output problems
Table 14-3: Milliamp output problems and recommended actions
Problem
Possible causes
Recommended actions
No mA output
•
Output not powered
•
Wiring problem
•
Circuit failure
•
Verify that the output loop is powered
externally.
•
Check the power supply and power supply
wiring. See
.
•
Verify the output wiring.
•
Check the
Fault Action
.
•
Contact customer service.
Loop test failed
•
Output not powered
•
Power supply problem
•
Wiring problem
•
Circuit failure
•
Verify that the output loop is powered
externally.
•
Check the power supply and power supply
wiring. See
.
•
Verify the output wiring.
•
Check the
Fault Action
.
•
Contact customer service.
mA output below
4 mA
•
Open in wiring
•
Bad output circuit
•
Process condition below LRV
•
LRV and URV are not set correctly
•
Fault condition if
Fault Action
is set to
Internal Zero or Downscale
•
Bad mA receiving device
•
Check your process conditions against the
values reported by the device.
•
Verify the receiving device, and the wiring
between the transmitter and the receiving
device.
•
Check the settings of
Upper Range Value
and
Lower Range Value
. See
Range Value and Upper Range Value
.
•
Check the
Fault Action
.
Troubleshooting
Configuration and Use Manual
March 2021
MMI-20020954
136