API tables supported by the API referral application
Set up temperature and pressure data for API referral using the display
The API referral application uses temperature and, optionally, pressure data in its calculations. You must
decide how to provide this data, then perform the required configuration and setup.
Tip
Fixed values for temperature or pressure are not recommended. Using a fixed temperature or pressure value
may produce inaccurate process data.
Prerequisites
The pressure measurement must be gauge pressure, not atmospheric pressure.
The pressure device must use the pressure unit that is configured in the transmitter.
If you are using an external temperature device, it must use the temperature unit that is configured in the
transmitter.
Procedure
1. Choose the method to be used to supply temperature data, and perform the required setup.
Method
Description
Setup
Internal
temperature
Temperature data from the on-
board temperature sensor
(RTD) will be used for all
measurements and
calculations. No external
temperature data will be
available.
a. Choose
Menu
>
Configuration
>
Process Measurement
>
Temperature
.
b. Set
External Temperature
to Off.
Configuration and Use Manual
Configure process measurement applications
MMI-20048166
April 2019
Configuration and Use Manual
45