Option
Description
Temperature-corrected
volume flow
The totalizer or inventory will track Temperature-Corrected Volume Flow
Rate and calculate total volume since the last reset.
Standard volume flow
The totalizer or inventory will track Standard Volume Flow Rate and
calculate total volume since the last reset.
Net mass flow
The totalizer or inventory will track Net Mass Flow Rate and calculate total
mass since the last reset.
Net volume flow
The totalizer or inventory will track Net Volume Flow Rate and calculate
total volume since the last reset.
Note
The totalizer/inventory value will not automatically be reset when the source is changed. The user must
manually reset the totalizer/inventory.
Tip
If you are using the API referral application and you want to measure batch-weighted average density
or batch-weighted average temperature, you must have a totalizer configured to measure
temperature-corrected volume flow.
3. Set
Totalizer Direction
to specify how the totalizer or inventory will respond to forward or reverse flow.
Option
Flow direction
Totalizer and inventory behavior
Forward Only
Forward
Totals increment
Reverse
Totals do not change
Reverse Only
Forward
Totals do not change
Reverse
Totals increment
Bidirectional
Forward
Totals increment
Reverse
Totals decrement
Absolute Value
Forward
Totals increment
Reverse
Totals increment
Important
Actual flow direction interacts with
Sensor Flow Direction Arrow
to determine the flow direction that
the transmitter uses in processing. See the following table.
Table 6-1: Interaction between actual flow direction and Sensor Flow Direction Arrow
Actual flow direction
Setting of Sensor Flow Direction
Arrow
Flow direction sent to outputs
and totalizers
Forward (same direction as Flow
arrow on sensor)
With Arrow
Forward
Against Arrow
Reverse
Reverse (opposite from Flow arrow
on sensor)
With Arrow
Reverse
Configure advanced options for process measurement
Configuration and Use Manual
April 2019
MMI-20048166
74