Reference Manual
00809-0100-4793, Rev CA
August 2004
5-5
Rosemount 8742C
1.
Right-click on the Transducer400 icon to initiate the empty pipe
learning function. This will bring up the Transducer block menu items.
From this menu, select Calibrate-Learn Empty Pipe, as shown in the
diagram below.
2.
The initial screen describes the process to be performed and any
configuration requirements. The entire two-minute process will be
performed. This is the last opportunity to cancel the opportunity and
return to normal status. To start the process, click Next.
3.
The transmitter will be put in Out of Service mode. The coil drive will
be set to 5 Hz. The learning mode in 5 Hz mode is a 60 second
process. A countdown clock is shown in the window. No intervention
is required.
4.
The transmitter will then automatically change to 37.5 Hz coil drive
frequency and repeat the same process as in the 5 Hz coil drive
mode.
5.
After both sets of data are written to the device, a check is made on
both the coil frequency and the mode. If the starting coil frequency
was 5 Hz, the method will return it to 5 Hz. Likewise, if the starting
mode was not Out of Service, the method will return the device to
Auto mode. Clicking
Finish
from the final screen will exit the method.
Содержание 8700 Series
Страница 2: ......
Страница 4: ......
Страница 12: ...Reference Manual 00809 0100 4793 Rev CA August 2004 1 4 Rosemount 8742C ...
Страница 64: ...Reference Manual 00809 0100 4793 Rev CA August 2004 5 12 Rosemount 8742C ...
Страница 102: ...Reference Manual 00809 0100 4793 Rev CA August 2004 Rosemount 8742C A 32 ...
Страница 116: ...Reference Manual 00809 0100 4793 Rev CA August 2004 Rosemount 8742C B 14 ...
Страница 135: ......