Emerson Process Management GmbH & Co. OHG
3-1
X-STREAM XEFD
Instruction Manual
HASXEDE-IM-EX
07/2017
3
Startup
Chapter 3
Start up
OPERATION AT LOW TEMPERATURES
When operating an instrument at temperatures below 0 °C (32 °F), do
NOT apply gas nor operate the internal pump before the warmup time has
elapsed!
Violation may result in condensation inside the gas paths or damaged
pump diaphragm!
EXPLOSION HAZARD
Ensure all covers, plugs and housing parts are in place and secured properly
before supplying power and signal voltages!
Ensure all requirements given within this manual for performing gas analysis
within a flameproof enclosure are considered BEFORE supplying gases
!
3.1 Final Check
Ensure that the analyzer has been installed
according to the descriptions in Chapter 2,
and that all covers and doors are closed and
fastened.
FLAMMABLE GASES – EXPLOSION HAZARD
Leaks may cause explosion when measuring flammable gases!
When measuring flammable gases it is recommended to perform a leak test
on all gas paths, connections and components before startup or applying
power. Leak tests should be carried out on a 2 month’s regular basis and
after repair/maintenance.