Emerson Process Management GmbH & Co. OHG
S-9
X-STREAM XEFD
Instruction Manual
HASXEDE-IM-EX
07/2017
S
Safety Instructions
Safety Instructions
EXPLOSION and ELECTRICAL SHOCK HAZARD BY INSTALLATION
These instruments provide a protective earth terminal. To prevent electrical
shock and explosion hazards, the instrument must be connected to a
protective earth.
Therefore the instrument has to be connected to mains by using a three
wire mains cable with earth conductor!
Any interruption of the earth connector inside or outside the instrument
or disconnecting the earth terminal may cause potential electrical shock
hazard!
Intended interruption of protective earth connections is not permitted!
EXPLOSION HAZARD BY DAMAGED FLAME PATHS
Take care to not damage the housing nor threads, and to not produce
scratches on the flange, as threads and flange function as flame paths!
Violation may result in explosion and personal injury!
Repair of the flamepaths is not intended.
ELECTRICAL SHOCK HAZARD WHEN CONNECTING TO MAINS
Before connecting the analyzer to mains power, please read the chapter on
safety warnings and the following instructions carefully.