24
Temperature Probe
To calibrate the temperature probe select “Calibration Temp” from the “Probe” menu. In order to complete this
task, it is necessary to use a professional thermometer to take the temperature of the plant.
Warning: this procedure can be completed only if the instrument is correctly installed and the
temperature probe is in good condition. Otherwise you could get unreliable results.
Take the temperature of the plant using a thermometer and enter it in the “Calib at.” field. Then press the knob
and move the cursor on “OK”. Confirm by pressing the knob again. End by moving the cursor to “EXIT” and
pressing the knob. Then turn the knob to select “YES” if you want to save the changes or “NO” to exit, restoring
the previous values. Then press the knob again to enforce your choice.
Temperature Alarm
Use this function to set a temperature value above which the instrument stops all activities, activates the main
alarm output and generates a warning message. To set the temperature value proceed as follows:
- press on the knob
- set the value by turning the knob
- press the knob and move the cursor on "ESC". Then press the knob again to confirm ("YES") and save the
setting.
pH probe calibration menu
Temperature calibration menu
ECL probe selection
ORP probe calibration menu