18.0 – ELECTRODE CLEANING AND MAINTENANCE 

The formation of deposits on the electrode will produce erroneous readings. The necessary cleaning action 
depends  on  the  type  of  deposit  under  consideration.  In  case  of thin deposits,  shake  the  electrode  or  use  a 
spray  of  distilled  water.  Organic  residues  or  parti cularly  resistant  deposits  should  be  removed  by  chemical 
means.  Mechanical  cleaning  of  the  bulb  should  be  resorted to only in  extreme  cases,  but bear in  mind  that 
abrasion can lead to irreparable damage. If cleaning does not fully re-establish the efficacy of the electrode, it 
may be that the electrode has aged. Ageing will show itself in the for of either a measurement error or a slow 
response. 

RECONDITIONING  

The following chemical treatments may be carried out with a view to reconditioning the electrode. We list them 
in order of increasing aggressiveness a s far as the bulb is concerned and would ask you to note that they will 
not necessarily improve the electrode quality, which in some cases may even be further reduced. 
1) Immerse the tip of the electrode in 0.1N hydrochloric acid (HCl) for 15 seconds, then rinse with water and 
dip again the electrode in a 0.1N solution of sodium hydroxide (NaOH) for 15 seconds, followed by a second 
rinsing.  Repeat  this  sequence  three  times,  and  then  carry  out another  check  measurement.  If  the  reading 
remains erroneous, proceed to treatment 2. 
2) Immerse the tip of the electrode in a 20% soluti on of ammonium bifluoride (NH2F-HF) for two or three 
minutes, then rinse with water and make another check measurement. If the reading is still unacceptabl e, 
proceed to treatment 3. 
3) Immerse the tip of the electrode in 5% hydrofluoric acid (HF) for 10 seconds, then rinse thoroughly in water 
and very quickly in 5N hydrochloric acid (HCl), followed by a second rinsing in water. If the check 
measurement still produces unacceptable measurements, it only remains to change the electrode. 
 

17.1 - ELECTRODE HOLDERS

 

Different types  of  electrode  holders  are  available:  for immersion,  electrode  cell, and  in-line. It is  essential  to 
bear in  mind  that  the  distance between  the injection  point  and  the  electrode  must  never  be less  than  one 
meter. When this is not possible, an appropriate chemical mixer must be inserted between the injection point 
and the electrode. 
 

EMAUX ELECTRODE HOLDERS FOR  CTRL2-7L APPLICATIONS

 

 

 

 ECELL1

                                                                                                   

ECELL 4

 

 ELINE 1 / ELINE 1-A 

    ELINE 2 

     ESUB 

 

 
 
 

Part no. 

Model  

SPS0000901 

ECELL1: Off-line holder 1 electrode, earth external valve for incoming flow control  

SPS0001201  

ECELL4: Off-line electrode and temperature probe holder, with Proximity switch 

2143002 

ESUB 0,5 m submersible holder, PVC body, length 50 cm 

2143003 / 2143004 

ESUB 1 m / 1,5 m: submersible holder, PVC body, length 100 cm / 150 cm 

2143001  

ELINE1 (PVC) In-line holder, PVC, pg3,5 1/2” 

DPS0001601 

ELINE1-A (AISI) In-line holder, AISI 316L, pg3,5 1/2”  

1PRS005 

ELINE2 (PP) In-line holder, pg3,5 1/2” 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Содержание CTRL2-5L

Страница 1: ...CTRL2 5L CTRL2 7L SERIES DUAL PARAMETER PH RX MICROCONTROLLER SOLENOID DRIVEN DOSING PUMP OPERATING INSTRUCTIONS AND MAINTENANCE EMPU12053149 ENGLISH...

Страница 2: ...ARAMETERS AND FUNCTIONS LAYOUT pH default page 12 10 0 PROGRAMMING CONTROLLER 11 0 CALIBRATION page 13 15 12 0 PROGRAMMING SET POINT 13 0 SET POINT SETTINGS page 15 17 14 0 ALLARM SETTINGS 15 0 RESET...

Страница 3: ...ger However save packaging it can be used for future shipments Before connecting the equipment ensure voltage ratings corresponds to local power supply You will find these characteristics in the produ...

Страница 4: ...IALS PARTS IN CONTACT WITH LIQUID STANDARD VERSION UPON REQUEST Pump head PP PVDF PMMA MEMO1 NOTE AISI316 Diaphragm PTFE Ball check valve CERAMIC AISI 316 Pump head Nipples Hose nut PP PVDF AISI 316 S...

Страница 5: ...nt side for sliding pump rear slots Mounting bracket side view Fig 3 Drill two holes and insert dibbles then screw in bracket Slide pump into slots Performance curves fig 2 are obtained testing the pu...

Страница 6: ...7 always use curved joints with suction or discharge hose Try to keep both the suction and discharge hose as straight as possible avoiding all unnecessary bends always use our standard strainer foot v...

Страница 7: ...e condition should be checked at regular intervals because excessive wear could cause additive to drip into the plant even when the pump is OFF If the problem persists install a proper calibrated coun...

Страница 8: ...ls are pumped especially If a reduction of the required additive concentration in the process plant is detected this could be caused the reduction of the dosing pump flow rate by valves wearing in whi...

Страница 9: ...ns CONNECTOR 1 CONNECTOR 1 Level switch connection fig 15 Configurations Pin 1 NO connection Pin 2 NO connection Pin 3 Level switch wire Pin 4 Level switch wire PUMP LOWER SIDE Power cable clamp pH or...

Страница 10: ...al mode selection of proportional range value Delay adjustment Delay at powering ON electrodespolarization Stand by for cleaning electrodesand sensors or re calibration 7 3 LEVEL CONTROL ONLY CTRL2 L...

Страница 11: ...quency based to maximum measuring value 9 0 PARAMETERS AND FUNCTIONS pH default DEFAULT Set point setting 7 2 Hysteresi s setting 0 1 Choice of intervention of dosing field acid alkaline Acid Selectio...

Страница 12: ...ogramming moves to measuring mode display will show current measured value Press OK to enter menus it will show SETPOINT menu using or buttons to scroll all 4 menus Set point menu Calibration menu Ala...

Страница 13: ...ature compensation that is setting the temperature at the same value of the system to be treated and such value will remain constant Also it s possible to switch from C celsius defau lt to F Fahrenhei...

Страница 14: ...POINT 2 Dip RX electrode into buffer solution 475 or 250 or 650mV Using or adjust until display shows the buffer solution value Press OK to confirm BNC CONNECTOR 12 0 PROGRAMMING SET POINT NOTE To ha...

Страница 15: ...ed by the user at max frequency chapter 11 5 and will stop once reached required set point PROP with this mode the pump will dose proportionally to the current measured value and its distance to selec...

Страница 16: ...00 120 imp min STEP 3 With and select minimum measured value to correspond dosing pump minimum injection frequency Press OK to confirm and exit Set point menu Press ESC to return to measuring mode 15...

Страница 17: ...are steps for RESET procedure Switch OFF and ON the unit Press at the same time ESC and OK until display will show the latest firmware revision e g 02 3 3 then will appear SET UP and default paramete...

Страница 18: ...ion d Check system pressure cause 2 chemical can produce crystalline deposits in the hydraulic parts or clogging the filter 2 GREEN LED LIT AND RED LED FLASHING PUMP PULSES BUT PUMP IS NOT PRIMING NO...

Страница 19: ...t manufacturer Customer Service Dealer or Distributor 8 LIQUID IS LEAKING from back of the pump head cause 1 diaphragm pump head o ring is not sealing because worned out or faulty due to not compatibl...

Страница 20: ...If the reading remains erroneous proceed to treatment 2 2 Immerse the tip of the electrode in a 20 solution of ammonium bifluoride NH2F HF for two or three minutes then rinse with water and make anoth...

Страница 21: ...njection fitting Injection non return valve N Delivery line to process plant O By pass from system line to electrode cell holder P ECELL1 off line electrode holder code SPS0000901 Q PH or RX electrode...

Страница 22: ...EMAUX customer service 1 pump head complete 2 diaphragm pump head o ring 3 PTFE diaphragm 4 piston flange 5 9 power cable assembly 6 front controls panel 7 housing 8 controls PCB printed circuit boar...

Страница 23: ...ir bleed 5 VALVE SPACER air bleed 6 P HEAD BALL CHECK VALVE ASSEMBLY 7 NIPPLE O RING 8 NIPPLE PP 3 8 9 PROTECTION CAP 10 HOSE NUT PP 11 12 13 14 15 AIR BLEED KNOB ASSEMBLY 16 17 PUMP HEAD SCREWS 12 18...

Страница 24: ...HOSE NUT PVDF 8 DIAPHRAGM PUMP HEAD O RING 9 PUMP HEAD SCREWS BUSH 10 DISCHARGE BALL CHECK ASSEMBLY 11 AIR BLEED BALL CHECK ASSEMBLY 12 AIR BLEED KNOB ASSEMBLY 13 O RING BLEED OUTLET 14 PUMP HEAD SCR...

Страница 25: ...RING 9 PUMP HEAD SCREWS BUSH 10 11 AIR BLEED BALL CHECK ASSEMBLY 12 DISCHARGE BALL CHECK ASSEMBLY 13 PUMP HEAD SCREWS Assembly includes Housing and Nipple 3 8 in PVDF PVDF PUMP HEAD LIP VALVE 2 VALVE...

Страница 26: ...ON CAP 10 HOSE NUT 11 DIAPHRAGM PUMP HEAD O RING 12 PUMP HEAD SCREWS 13 PUMP HEAD SCREWS BUSH ONLY CTRL2 7L SERIES OVERALL DIMENSIONS mm PMMA pump head polymer viscous liquid configuration includes Ce...

Страница 27: ...alves in FPM standard EPDM SILICON NBR INJECTION VALVE FITTING NO RETURN VALVE STANDARD INJECTION VALVE A CYLINDER SLEEVE NO RETURN VALVE B NIPPLE O RING 2062 C INJECTION FITING PP D HOSE NUT PP Valid...

Страница 28: ...cables of the previous mentioned components A SUCTION LANCE B LEVEL SWITCH C FOOT VALVE FILTER D AIR BLEED HOSE E PUMP SUCTION HOSE F LEVEL SWITCH CABLE Suction lance is made in PVC and are available...

Страница 29: ...in PP PVC or PVDF RELIEF VALVE or BACK PRESSURE VALVE This valve can be utilized either as RELIEF VALVE opens an overflow line if the preset pressure limit is exceeded This application also is a safet...

Страница 30: ...30 MEASUREMENTS AND PROGRAMMING NOTES...

Страница 31: ...RL2 Version Emaux Zhong Shan Swimming Pool Equipment Co Ltd Long Zhu Industrial Park Nan Lang Industrial Area Nam Lang Town ZhongShan City Guangdong Province China Tel 0760 8552 7988 Fax 0760 8552 718...

Отзывы: