emaux CTRL2-5L Скачать руководство пользователя страница 18

 

1

8

17.0 – TROUBLE-SHOOTING FOR CTRL2 SERIES 

Operator must always protect itself w hen handling hazardous feed chemicals. Use googles and 

protectiv e clothing and follow the safety information sheets of chemical manufacturer. 

 When remov ing the metering pump from the plant, be careful as there might be some residual 

additiv e in the discharge hose. 
- Due to robust design the pump is quite robust therefore there are no apparent mechanical problems. 
- In case of leakage from nipples, check tube nut tight setting. 
- Diaphragm rarely breaks (made of pure PTFE synthesized), however check diaphragm-pump head o-ring. 
-  Occasionally  there might  be a  loss  of  liquid  from  the  nipple because  the tube  nut  has  loosened, or  more 

simply  the  discharge  hose  has  broken.  Very  rarely  there  may  be  losse s  caused  by  the  breakage  of  the 
membrane, or  by  the membrane  seals  in  which  case  they  have  to  be  replaced  by  disassembling  the  four 
screws of the pump head, when re-mounting the pump head ensure that the screws are replaced properly, 
properly, along with “O” ring. 

- re-tighten liquid end screws after 24 hours in operation. 
- After repair, metering pump will need to be cleaned of additive residues which can damage pump casing. 

 

TROUBLE-SHOOTING QUICK VIEW 

 

PROBLEM 

CAUSE 

ACTION TO TAKE: 

1

 

 

GREEN LED LIT 

AND RED LED 
FLASHING,

 

DOSING 

PUMP PULSES BUT 
ADDITIVE IS NOT 
INJECTED 
 

cause 1: diaphragm rupture and/or 
pump head o-ring is not sealing 
because worn out or not compatible 
with liquid. 

a.  Dismount  the  suction  and  discharge 
valves,  clean  them  reinstall  them  or  replace 
(pump head views in the following pages). In 
case of using LIP type valves and should the 
valves  be  swollen,  check  valves  material 
against our chemical resistance compatibility 
chart and fit correct valves.  
b. Check clogging of the filter 
c. Check injection valve condition  
d. Check system pressure 

cause 2: chemical can produce 
crystalline deposits in the hydraulic 
parts or clogging the filter. 

2

 

 

GREEN LED LIT 

AND RED LED 
FLASHING,

 

PUMP 

PULSES BUT PUMP 
IS NOT PRIMING (NO 
LIQUID SUCKED) 

a. Dismount the suction and discharge valves, clean them and replace, see position 
(fig.10) and follow the same point as above point 1. 
b. In case of priming difficulties: use a normal syringe to suck liquid from the 
discharge nipple while the pump is in operation (ai r bleed valve closed), continuing 
until you actually see the liquid rise in the syringe. Use a short piece of suction PVC 
hose to connect  the syringe to the discharge nipple (see photo sequence below)
 

cause2: fuse blown or defective.  Also a 

possible voltage spike higher then pump 

range has blown fuse or damaged pcb.

Disconnect the pump from power supply, check the 

fuse and if blown or defective change it with same 

type or contact manu facturer Customer Service, 

Dealeror Distributor. CTRL2 series are equipped 

with a 

MICROFUSE MT 1A slow.

DO NOT REPLACE FUSE WITH OTHER TYPES: it will severely damage pump and other equipment !!!

In case of severe and continuous voltage fluctuation, the fuse could blow continuously. In this case contact

manufacturer Customer Service, Dealer or Distributor for a correct solution.

 

Insert syringe hose in 
place of discharge hose 
ensuring that the same 
is NOT connected to 
injection valve. 

 

Close pump head air bleed valve. 
Start pump pulses above 50% rate and start sucking with 
syringe until the liquid start to come up costantly in syringe 
hose then stop the pump and take out syringe hose. 

 

Reinsert discharge 
hose ensuring that the 
same is connected to 
injection valve and 
start the pump. 

3

 

 ALL LED 

DISPLAYS ARE OFF, 
THE PUMP DOES 
NOT PULSE 
 

cause 1: power supply cable not 
plugged or wrong  power supply or 
plug not wired correctly. 

Check  power  supply  (socket,  plug,  power 
switch  ON), if the  pump  doesn’t  work  contact 
manufacturer  Customer  Service,  Dealer  or 
Distributor. Ensure system is well grounded,  

 

Содержание CTRL2-5L

Страница 1: ...CTRL2 5L CTRL2 7L SERIES DUAL PARAMETER PH RX MICROCONTROLLER SOLENOID DRIVEN DOSING PUMP OPERATING INSTRUCTIONS AND MAINTENANCE EMPU12053149 ENGLISH...

Страница 2: ...ARAMETERS AND FUNCTIONS LAYOUT pH default page 12 10 0 PROGRAMMING CONTROLLER 11 0 CALIBRATION page 13 15 12 0 PROGRAMMING SET POINT 13 0 SET POINT SETTINGS page 15 17 14 0 ALLARM SETTINGS 15 0 RESET...

Страница 3: ...ger However save packaging it can be used for future shipments Before connecting the equipment ensure voltage ratings corresponds to local power supply You will find these characteristics in the produ...

Страница 4: ...IALS PARTS IN CONTACT WITH LIQUID STANDARD VERSION UPON REQUEST Pump head PP PVDF PMMA MEMO1 NOTE AISI316 Diaphragm PTFE Ball check valve CERAMIC AISI 316 Pump head Nipples Hose nut PP PVDF AISI 316 S...

Страница 5: ...nt side for sliding pump rear slots Mounting bracket side view Fig 3 Drill two holes and insert dibbles then screw in bracket Slide pump into slots Performance curves fig 2 are obtained testing the pu...

Страница 6: ...7 always use curved joints with suction or discharge hose Try to keep both the suction and discharge hose as straight as possible avoiding all unnecessary bends always use our standard strainer foot v...

Страница 7: ...e condition should be checked at regular intervals because excessive wear could cause additive to drip into the plant even when the pump is OFF If the problem persists install a proper calibrated coun...

Страница 8: ...ls are pumped especially If a reduction of the required additive concentration in the process plant is detected this could be caused the reduction of the dosing pump flow rate by valves wearing in whi...

Страница 9: ...ns CONNECTOR 1 CONNECTOR 1 Level switch connection fig 15 Configurations Pin 1 NO connection Pin 2 NO connection Pin 3 Level switch wire Pin 4 Level switch wire PUMP LOWER SIDE Power cable clamp pH or...

Страница 10: ...al mode selection of proportional range value Delay adjustment Delay at powering ON electrodespolarization Stand by for cleaning electrodesand sensors or re calibration 7 3 LEVEL CONTROL ONLY CTRL2 L...

Страница 11: ...quency based to maximum measuring value 9 0 PARAMETERS AND FUNCTIONS pH default DEFAULT Set point setting 7 2 Hysteresi s setting 0 1 Choice of intervention of dosing field acid alkaline Acid Selectio...

Страница 12: ...ogramming moves to measuring mode display will show current measured value Press OK to enter menus it will show SETPOINT menu using or buttons to scroll all 4 menus Set point menu Calibration menu Ala...

Страница 13: ...ature compensation that is setting the temperature at the same value of the system to be treated and such value will remain constant Also it s possible to switch from C celsius defau lt to F Fahrenhei...

Страница 14: ...POINT 2 Dip RX electrode into buffer solution 475 or 250 or 650mV Using or adjust until display shows the buffer solution value Press OK to confirm BNC CONNECTOR 12 0 PROGRAMMING SET POINT NOTE To ha...

Страница 15: ...ed by the user at max frequency chapter 11 5 and will stop once reached required set point PROP with this mode the pump will dose proportionally to the current measured value and its distance to selec...

Страница 16: ...00 120 imp min STEP 3 With and select minimum measured value to correspond dosing pump minimum injection frequency Press OK to confirm and exit Set point menu Press ESC to return to measuring mode 15...

Страница 17: ...are steps for RESET procedure Switch OFF and ON the unit Press at the same time ESC and OK until display will show the latest firmware revision e g 02 3 3 then will appear SET UP and default paramete...

Страница 18: ...ion d Check system pressure cause 2 chemical can produce crystalline deposits in the hydraulic parts or clogging the filter 2 GREEN LED LIT AND RED LED FLASHING PUMP PULSES BUT PUMP IS NOT PRIMING NO...

Страница 19: ...t manufacturer Customer Service Dealer or Distributor 8 LIQUID IS LEAKING from back of the pump head cause 1 diaphragm pump head o ring is not sealing because worned out or faulty due to not compatibl...

Страница 20: ...If the reading remains erroneous proceed to treatment 2 2 Immerse the tip of the electrode in a 20 solution of ammonium bifluoride NH2F HF for two or three minutes then rinse with water and make anoth...

Страница 21: ...njection fitting Injection non return valve N Delivery line to process plant O By pass from system line to electrode cell holder P ECELL1 off line electrode holder code SPS0000901 Q PH or RX electrode...

Страница 22: ...EMAUX customer service 1 pump head complete 2 diaphragm pump head o ring 3 PTFE diaphragm 4 piston flange 5 9 power cable assembly 6 front controls panel 7 housing 8 controls PCB printed circuit boar...

Страница 23: ...ir bleed 5 VALVE SPACER air bleed 6 P HEAD BALL CHECK VALVE ASSEMBLY 7 NIPPLE O RING 8 NIPPLE PP 3 8 9 PROTECTION CAP 10 HOSE NUT PP 11 12 13 14 15 AIR BLEED KNOB ASSEMBLY 16 17 PUMP HEAD SCREWS 12 18...

Страница 24: ...HOSE NUT PVDF 8 DIAPHRAGM PUMP HEAD O RING 9 PUMP HEAD SCREWS BUSH 10 DISCHARGE BALL CHECK ASSEMBLY 11 AIR BLEED BALL CHECK ASSEMBLY 12 AIR BLEED KNOB ASSEMBLY 13 O RING BLEED OUTLET 14 PUMP HEAD SCR...

Страница 25: ...RING 9 PUMP HEAD SCREWS BUSH 10 11 AIR BLEED BALL CHECK ASSEMBLY 12 DISCHARGE BALL CHECK ASSEMBLY 13 PUMP HEAD SCREWS Assembly includes Housing and Nipple 3 8 in PVDF PVDF PUMP HEAD LIP VALVE 2 VALVE...

Страница 26: ...ON CAP 10 HOSE NUT 11 DIAPHRAGM PUMP HEAD O RING 12 PUMP HEAD SCREWS 13 PUMP HEAD SCREWS BUSH ONLY CTRL2 7L SERIES OVERALL DIMENSIONS mm PMMA pump head polymer viscous liquid configuration includes Ce...

Страница 27: ...alves in FPM standard EPDM SILICON NBR INJECTION VALVE FITTING NO RETURN VALVE STANDARD INJECTION VALVE A CYLINDER SLEEVE NO RETURN VALVE B NIPPLE O RING 2062 C INJECTION FITING PP D HOSE NUT PP Valid...

Страница 28: ...cables of the previous mentioned components A SUCTION LANCE B LEVEL SWITCH C FOOT VALVE FILTER D AIR BLEED HOSE E PUMP SUCTION HOSE F LEVEL SWITCH CABLE Suction lance is made in PVC and are available...

Страница 29: ...in PP PVC or PVDF RELIEF VALVE or BACK PRESSURE VALVE This valve can be utilized either as RELIEF VALVE opens an overflow line if the preset pressure limit is exceeded This application also is a safet...

Страница 30: ...30 MEASUREMENTS AND PROGRAMMING NOTES...

Страница 31: ...RL2 Version Emaux Zhong Shan Swimming Pool Equipment Co Ltd Long Zhu Industrial Park Nan Lang Industrial Area Nam Lang Town ZhongShan City Guangdong Province China Tel 0760 8552 7988 Fax 0760 8552 718...

Отзывы: