Servicing
51
11.1 Starting non-stop operation
The automatic repeating of complete cycles is possible as non-stop operation without
combustion.
The Autoloader keeps repeating all individual steps until the last crucible has been removed
from the crucible tray. If no further crucible is found, the Autoloader continues to turn the chain
for a while to look for more crucibles. If no more crucibles are found, the chain is stopped. The
non-stop operation is ended.
The following conditions must be met:
•
Commissioning must have been completed, see
”.
Start non-stop operation as follows:
1. Completely load the Autoloader with empty crucibles.
2. A
ctivate manual mode, see “
Activating manual operating mode
”.
3. Secure the button for manual control using an elastic ring.
4. Set the mains switch on the analyzer from Position 0 to Position 1.
Non-stop operation has been started.