Installation
36
Manually run through the motion sequences as follows:
1. Ensure that the assembly steps 5.1 to 5.7 have been carried out.
2. Ensure that several empty crucibles are ready on the crucible tray.
3. Ensure that the crucible bin has been correctly positioned.
4. Press the button (
2
) for manual control. The first step in the motion sequence is carried
out. The crucible chain conveys the first crucible to the pick-up position. Each additional
press of the button performs the next step in the Autoloader work sequence.
5. Align the Autoloader so that the gripper arm can place the crucible onto the middle of the
crucible tray on the analyzer and pick it up again there. To do this, move the Autoloader.
6. Set the adjusting screws on the four Autoloader bases (
1
) as desired to adjust the height of
the Autoloader. The height has been correctly set when there is a small gap between the
crucible in the gripper arm and the crucible tray on the analyzer before opening the gripper
fingers.
Additionally, for the 130 autoloader: Align the autoloader horizontally across the width
using a spirit level. Use the two right feet (
5
) for this.
7. Repeat steps 4. to 6. of the motion sequence several times until the crucible is positioned
securely by the Autoloader in the middle of the crucible tray and picked up again.
8. Now tighten all loose screws (
3
) and (
4
).
9. Disconnect the power cord on the power supply unit from the Autoloader.
The Autoloader has been aligned.