93
-
Shift the switch with holder downwards until the magnetic switch is off (A
valve clicks and the piston slightly moves downwards).
-
Make a second marking on the cylinder (6) at the upper end of the switch
holder (4).
-
Adjust the switch holder (4) to have the upper end in the middle of the two
markings.
-
Open the furnace and then close it again (F2) to verify the correctness of the
adjustment.
-
Adjust the lower cushioning (7), so that the furnace opens without the piston
banging at the bottom.
-
Click on “Communication test”-“Digital outputs”, on the 9th button counting
from the left, to energize the coil cooling valve (11/V8), or start analysis
cycle.
-
Turn the restrictor (12) counter clockwise until the pressure on the gauge
(13) is reduced by about 0.2 to 0.3 bar (3 to 4 psi) when the coil cooling
valve (11/V8) is open.
-
Fix the locknut of the restrictor (12).
Remark:
In case of using valuable gas for operating the pneumatics, the restrictor (12)
should be adjusted for lower flow by turning it clockwise. It can even be completely
closed by turning it clockwise up to the end of its range.
The flow adjusted with this restrictor is for cooling the combustion coil to ensure
very long service life. According to our experience, the combustion coils of our
analyzers don’t need replacement, even after years of operation.
Pos : 7. 19 /0020 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Press ure regul ati on ELTR A @ 6\ mod_1369298860002_9. doc x @ 41930 @ 32233 @ 1
4.4
46B
Pressure regulation
Pos : 7. 20 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-800_Ser vic e/ 0020 Eins tell ungen, Tests und F unktions beschr eibung/0255 M odul Pres sur e R egulation @ 6\ mod_1377688630939_9.doc x @ 46975 @ 2424 @ 1
Fig. 48: PR1
PR1 is the inlet oxygen regulator, and is situated inside the analyzer, on the rear
panel.
The outlet pressure needs to be set to 1.5bar (22.5psi).
•
Pull and rotate the knob of the regulator, until the pressure gauge (6) shows
the correct value.
–
Ensure that the flow shown on the flow meter (5) shows between 180 and
200l/h; if not, adjust according to the related chapter.
Содержание CS-800
Страница 2: ...2 Copyright Copyright by Eltra GmbH Haan Retsch Allee 1 5 D 42781 Haan Federal Republic of Germany ...
Страница 3: ...3 ...
Страница 10: ...10 Fig 2 Draft 12057 ...
Страница 11: ...11 Fig 3 Draft 11660 ...
Страница 12: ...12 Fig 4 Draft 12200 ...
Страница 14: ...14 Fig 6 Draft 12058 Signal and control ...
Страница 17: ...17 Fig 8 Draft 6 2 4 Gas flow diagram furnace closed ...
Страница 18: ...18 Fig 9 Oxygen flow schematically ...
Страница 19: ...19 Fig 10 Draft 12059 signal and control board ...
Страница 21: ...21 Fig 12 Draft VG 96 connector ...
Страница 22: ...22 Fig 13 Draft 1 14 4 1 ...
Страница 23: ...23 Fig 14 Draft IR 45 power supply ...
Страница 24: ...24 Fig 15 Draft IR sources Fig 16 Draft 7 pin plug ...
Страница 80: ...80 Abb 46 IRC 1 3 ...
Страница 82: ...82 Fig 29 Mainboard Cut out IR Cell general test points ...
Страница 91: ...91 Abb 47 Pneumatics parts 2 1 3 4 5 6 7 8 9 10 11 12 13 14 15 ...
Страница 94: ...94 bar 0 5 1 1 5 2 2 5 0 PR2 1 Fig 49 Front side 4 5 47BPurge pressure Fig 50 PR2 5 6 ...
Страница 109: ...Service 109 10 10 11 9 b7 10 1 12 8 5 6 Fig 60 b7 10 1 ...
Страница 110: ...Service 110 ...
Страница 119: ...Miscellaneous 119 77510 Heaters for IR cell ...
Страница 120: ...Miscellaneous 120 6 1 2 121BFront side Fig 41 Front side ...
Страница 122: ...Miscellaneous 122 6 1 3 122BRight side Fig 42 Right side ...
Страница 126: ...Miscellaneous 126 6 1 5 124BOscillating circuit Fig 44 Oscillating circuit ...
Страница 130: ...Miscellaneous 130 6 1 7 126BFurnace cleaning 75130 14080 75122 70120 14072 14051 14045 75150 14021 14160 ...
Страница 132: ...Miscellaneous 132 21010 Heather section 21120 Quartz tube 70380 O Ring 70380 21120 21010 ...
Страница 134: ...Miscellaneous 134 Fig 47 12055 I ...
Страница 135: ...Miscellaneous 135 Fig 48 12055 II ...
Страница 136: ...Miscellaneous 136 Fig 49 12072 ...
Страница 137: ...Miscellaneous 137 Fig 50 12064 ...
Страница 138: ...Miscellaneous 138 Fig 51 12550 Vacuum cleaner ...
Страница 139: ...Miscellaneous 139 Fig 52 12015 I ...
Страница 140: ...Miscellaneous 140 Fig 53 12015 II ...
Страница 141: ...Miscellaneous 141 Fig 54 12016 ...
Страница 143: ...Miscellaneous 143 6 2 1 127BUNI 1 ...
Страница 144: ...Miscellaneous 144 6 2 2 128BGas flow system ...
Страница 145: ...Miscellaneous 145 ...
Страница 146: ...Miscellaneous 146 6 2 3 129BValve controlling Fig 55 Valve controlling I 11001 6001 ...
Страница 155: ......
Страница 156: ......
Страница 157: ......
Страница 158: ...Copyright Copyright by Eltra GmbH Haan Retsch Allee 1 5 D 42781 Haan Federal Republic of Germany ...