62
3.19.1
97B
Reasons for instability:
3.19.1.1 Too low deviation value
One reason for the above message is that the deviation value in the configuration
window is too low, so that a higher value should be entered. The levels of noise of
all cells are not necessarily identical. Longer cells have normally higher noise than
shorter ones and the sensors themselves have noise level deviations between
exemplars.
3.19.1.2 Too high noise
One more reason could be exceptionally high noise. The base lines and the level
of their noise can be seen on the screen when clicking on the base lines icon (with
three waves).
The most common reasons for too high noise are:
–
The infrared sensor.
–
The pre-amplifier,
–
A fault on the circuit board IRC-1.x where x can be 1,2, or 3.
–
At least one IR source is defective.
–
The microcontroller board.
–
The power supply fluctuates.
In case of a module (rack) of more than one IR-cell, mostly only one of the base
lines has too high noise while the others have noise in normal range. In order to
localize whether the noise is created by the sensor, preamplifier or the
corresponding section of the IRC board, interchange the preamp connector of the
noisy range with another one on the IRC board. If the noise moves to the other
range on the PC screen (base line color), this means that the noise is created by
the sensor or preamplifier.
In this case replace first the sensor as the more probable reason and if this doesn't
help, replace the preamplifier.
NOTE
Don't forget to connect the interchanged connectors to the original position again.
•
If the noise is created by the IRC board, change it.
Remark: In such case, theoretically, the noise could be created by the
corresponding analog input of the microcontroller board however this possibility is
very unlike.
3.19.1.3 Too high drift
Another reason for report of instability is a too high drift of at least one cell, so that
the signal of one cell reaches the end of the set window (mV).
A drift can be due to temperature changes of the IR cell. It happens when the cell
temperature is still not stable after switching on the power. When switching on the
power after long break so that the cell is cold, allow the cell at least 30 minutes to
stabilize the temperature. This has to be taken into consideration when checking
the base line stability.
When starting the oxygen flow (power switch on position 2), the cell takes about 5
minutes to stabilize when restarting the oxygen flow after long time without flow
(but with uninterrupted power).
Pos : 5. 43 /0020 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 F urnac e opens wit h del ay af ter the end of anal ysis ELT RA @ 6\ mod_1368703238004_9.doc x @ 41143 @ 222 @ 1
3.20
33B
Furnace opens with delay after the end of analysis
Pos : 5. 44 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-800_Ser vic e/ 0015 F ehl er/ 019510 M odul F urnac e opens wit h del ay ELTRA @ 6\ mod_1377688610559_9. doc x @ 46735 @ 22444 @ 1
There can be two different reasons for this.
1.
One reason is that this delay is set in the software. The software provides
the possibility to set a time to delay the furnace opening after the end of
analysis. This is occasionally required for samples creating harmful oxides or
Содержание CS-800
Страница 2: ...2 Copyright Copyright by Eltra GmbH Haan Retsch Allee 1 5 D 42781 Haan Federal Republic of Germany ...
Страница 3: ...3 ...
Страница 10: ...10 Fig 2 Draft 12057 ...
Страница 11: ...11 Fig 3 Draft 11660 ...
Страница 12: ...12 Fig 4 Draft 12200 ...
Страница 14: ...14 Fig 6 Draft 12058 Signal and control ...
Страница 17: ...17 Fig 8 Draft 6 2 4 Gas flow diagram furnace closed ...
Страница 18: ...18 Fig 9 Oxygen flow schematically ...
Страница 19: ...19 Fig 10 Draft 12059 signal and control board ...
Страница 21: ...21 Fig 12 Draft VG 96 connector ...
Страница 22: ...22 Fig 13 Draft 1 14 4 1 ...
Страница 23: ...23 Fig 14 Draft IR 45 power supply ...
Страница 24: ...24 Fig 15 Draft IR sources Fig 16 Draft 7 pin plug ...
Страница 80: ...80 Abb 46 IRC 1 3 ...
Страница 82: ...82 Fig 29 Mainboard Cut out IR Cell general test points ...
Страница 91: ...91 Abb 47 Pneumatics parts 2 1 3 4 5 6 7 8 9 10 11 12 13 14 15 ...
Страница 94: ...94 bar 0 5 1 1 5 2 2 5 0 PR2 1 Fig 49 Front side 4 5 47BPurge pressure Fig 50 PR2 5 6 ...
Страница 109: ...Service 109 10 10 11 9 b7 10 1 12 8 5 6 Fig 60 b7 10 1 ...
Страница 110: ...Service 110 ...
Страница 119: ...Miscellaneous 119 77510 Heaters for IR cell ...
Страница 120: ...Miscellaneous 120 6 1 2 121BFront side Fig 41 Front side ...
Страница 122: ...Miscellaneous 122 6 1 3 122BRight side Fig 42 Right side ...
Страница 126: ...Miscellaneous 126 6 1 5 124BOscillating circuit Fig 44 Oscillating circuit ...
Страница 130: ...Miscellaneous 130 6 1 7 126BFurnace cleaning 75130 14080 75122 70120 14072 14051 14045 75150 14021 14160 ...
Страница 132: ...Miscellaneous 132 21010 Heather section 21120 Quartz tube 70380 O Ring 70380 21120 21010 ...
Страница 134: ...Miscellaneous 134 Fig 47 12055 I ...
Страница 135: ...Miscellaneous 135 Fig 48 12055 II ...
Страница 136: ...Miscellaneous 136 Fig 49 12072 ...
Страница 137: ...Miscellaneous 137 Fig 50 12064 ...
Страница 138: ...Miscellaneous 138 Fig 51 12550 Vacuum cleaner ...
Страница 139: ...Miscellaneous 139 Fig 52 12015 I ...
Страница 140: ...Miscellaneous 140 Fig 53 12015 II ...
Страница 141: ...Miscellaneous 141 Fig 54 12016 ...
Страница 143: ...Miscellaneous 143 6 2 1 127BUNI 1 ...
Страница 144: ...Miscellaneous 144 6 2 2 128BGas flow system ...
Страница 145: ...Miscellaneous 145 ...
Страница 146: ...Miscellaneous 146 6 2 3 129BValve controlling Fig 55 Valve controlling I 11001 6001 ...
Страница 155: ......
Страница 156: ......
Страница 157: ......
Страница 158: ...Copyright Copyright by Eltra GmbH Haan Retsch Allee 1 5 D 42781 Haan Federal Republic of Germany ...