Elsys AG
www.elsys-instruments.com
15
TraNET FE User Guide
4.2 Measurement Settings
4.2.1 Device
• Name:
Set the device name as visible in TranAX
• Description:
Additional information about the device helping to identify the device in
the
network.
• Server port:
TCP Port, should only be changed if it is in conflict with another
application in the network. Default 10010
4.2.2 Measurement Settings:
• BDF File Path:
select if BDF files are stored to the internal or the external USB drive.
• Nr. Of BDF Files:
Defines how many old heap files will be preserved when several
consec utive measurements were started. If set to 0, on each new
continuous or ECR measurement the last heap file gets overwritten.
• Auto Start at Power Up:
The device starts the measurement automatically with the last settings
when the device powers up.
• Start Autosequence at Power Up:
If any remote autosequence/MFC are stored on the device, enabling this
option will automatically start the autosequence after power up.
• Use Local Time:
if enabled, the absolute measurement time is taken from the device real
time clock (RTC), otherwise the client computer from which the
measurement is started acts as time reference.
For using PTP or GPS synchronization, option must be enabled!
4.2.3 Synchronization Settings
• SyncLink 2 extra Delay:
Additional delays for compensating daisy-chained SyncLink 2 connection.
• Pulse Per Second (PPS) Source:
Selects the source of the PPS signal for timebase synchronization:
- Ethernet (PTP) uses the PPS from the PTP hardware
- GPS (internal) uses the PPS from the integrated GPS receiver
(Optional)
- External Timebase Input for using any external PPS signal
connected to the Ext-IO input “External Timebase”.
Attention: For using the PPS synchronization,”PPS Sync” must be activated on
the TranAX Control Panel.