11
TraNET FE User Guide
You can’t connect as written above?
Someone has maybe already changed the IP Address already. Set the device to DHCP and AUTO
IP mode:
• Switch off the TraNET FE device, press and hold the button in the hole next to the LED
“Ready”, power-on the device and wait until the LED “Ready” blinks permanently.
Alternatively, the button can also be pressed during runtime for at least 4 seconds. LED
“Ready” will stop flashing (stays green or off). DHCP and Auto-IP will be set, after approx.
10 seconds. The green LED “Ready” starts flashing again. IP Settings now have changed to
DHCP and AUTO IP mode.
• Network settings on the computer now have to be set to DHCP too!
pass
Done
Connect to
Device
Device Found
TranAX search for
devices
UDP Broadcast on port
10020/10021
pass
IP address configuration
successful
Power On
TraNET request a
valid IP address from
the DHCP server
TraNET FE
Connection Flow
fail
TraNET set an Auto IP
in the range
169.254.0.0 -
169.254.255.255
Test if address is unique
in the net
fail
pass
UDP port blocked?
Check Firewall Settings
for Port UDP
10020/10021
no device
found
Check Windows
Settings, Auto IP
address enabled?
NO
Enable Auto IP in
Windows (DHCP)
Check Firewall Settings,
Allow TranAX to communicate
over the Network
fail
TCP IP port 10010
blocked?
TranAX network
communication blocked?
Condition:
1:1 connection or network with DHCP server