46
Assembly
, Connection and Putting into Operation
Ill.: 11
Switch for electrical isolation
Output DA2 is electrically isolated
(the slider on the switch is up)
Output DA1 is not electrically isolated
(the slider on the switch is down)
5.3 Putting into operation
5.3.1 Configuration of measurement parameters
If the DL230 is subject to calibration regulations, the
works described below should only be performed by
legally authorized individuals.
The necessary operating parameters can be adjusted using the free
configuration program "enSuite", which is available under
www.elster-instromet.com.
If the program is not available, the configuration can also be carried
out using the keyboard as described below (see chapter 5.3.1.2).
Before adjusting the measurement parameters via the
keyboard, please read chapter 6 to learn how the device
is operated.
5.3.1.1 Opening the calibration lock
The calibration lock is located at the back of the housing cover in the
form of a button, and this can be secured by means of an adhesive
label. This button must be pressed in order to adjust the values and
parameters protected by configuration regulations.
Ill.:
12
Содержание DL230
Страница 1: ...Operating manual Data Logger DL230 ...
Страница 2: ...Data Logger DL230 Operating manual 73022387 Issue b Issue date 11 11 2014 Software version from 1 01 ...
Страница 8: ......
Страница 47: ...Assembly Connection and Putting into Operation 47 Position of the key to open the calibration lock ...
Страница 105: ...Appendix 105 English manual 73 022 387 ...
Страница 106: ...106 Appendix 9 2 EC Declaration of Conformity ...
Страница 107: ...Appendix 107 9 3 ATEX Type Examination Certificate 9 3 1 Associated apparatus for Zone 0 1 ...
Страница 108: ...108 Appendix ...
Страница 109: ...Appendix 109 ...
Страница 110: ...110 Appendix ...