48
Assembly
, Connection and Putting into Operation
5.3.1.2 Adjusting the parameters for the pulse transducer of the
meter
If a pulse transducer is connected as per 5.2.1.1 then adjust the
input mode and the cp value as follows:
1. Setting the input mode:
Move the cursor to the "Serv." tab and via the path below to the
input mode "Md.E1":
Serv.
Input 1
Settings
Md.E1
Press the ENTER button
The set value flashes.
Repeatedly press one of the
or
arrow keys until the text
"Pulse Input" flashes.
Press the ENTER key to confirm the value setting. The input
value can be deleted by using the ESC key.
2. Setting the cp (pulse constant) value:
Move the cursor to cp value "cp.E1" in the same path
Press the ENTER
button. One of the value's digits flashes.
Move the cursor to the digits using the
or
arrow keys and
update them using the
or
arrow keys.
Press the ENTER key to confirm the value setting. The input
value can be deleted by using the ESC key.
The same process can also be used for the other inputs,
inputs 2 to
4
.
5.3.1.3 Activating encoder mode
If an encoder is connected as per 5.2.1.2 the encoder mode is
activated as follows:
Start the "Encoder Recognition" function using the DL230 keypad as
follows:
Move the cursor to the "Serv." register and to the value "Md.I1"
(input mode) via the following path:
Serv.
Input 1
Parameter settings
Md.I1
Press the ENTER button
The set value flashes.
Repeatedly press one of the
or
arrow keys until the value
"Auto-Encoder" flashes.
Press the ENTER key to confirm the value setting. Input can be
cancelled using the ESC key.
Содержание DL230
Страница 1: ...Operating manual Data Logger DL230 ...
Страница 2: ...Data Logger DL230 Operating manual 73022387 Issue b Issue date 11 11 2014 Software version from 1 01 ...
Страница 8: ......
Страница 47: ...Assembly Connection and Putting into Operation 47 Position of the key to open the calibration lock ...
Страница 105: ...Appendix 105 English manual 73 022 387 ...
Страница 106: ...106 Appendix 9 2 EC Declaration of Conformity ...
Страница 107: ...Appendix 107 9 3 ATEX Type Examination Certificate 9 3 1 Associated apparatus for Zone 0 1 ...
Страница 108: ...108 Appendix ...
Страница 109: ...Appendix 109 ...
Страница 110: ...110 Appendix ...