SO250-UI Tank Sensor | Version 18.08.2015 | Data subject to alteration. Errors subject to correction.
5
The measuring track is aligned vertical to the sensor surface. The measuring track has
to meet the measuring surface vertically, too.
To avoid wave formation in fluids, the measured area can be separated using a
measuring tube (Ø at least 50 mm). Please note: the measuring area is only calm as
long as the end of the tube is below the surface of the liquid.
Area must not be hermetically
sealed so that the fluid level in
the tube may change.
Loud surrounding noise (e.g. when filling metal tanks) may disrupt the measurement.
Please contact us with any questions regarding the area of application or installation.
90°
90°
Sensor
Measuring
track
Measuring surface