Use the and keys to set the operating range of
the current output. Settings options: 0 mA … 20 mA
or 4 mA … 20 mA.
Confirm your selection using the key. You will automatically return to the main menu.
Operating range 0…20 mA:
Distance measurement:
Content measurement:
0 mA
12 cm
0 mA
0% Tank content
20 mA
250 cm
20 mA
100% Tank content
If "No echo received“: 0 mA.
Operating range 4…20 mA:
Distance measurement:
Content measurement:
4 mA
12 cm
4 mA
0% Tank content
20 mA
250 cm
20 mA
100% Tank content
If "No echo received“: 4 mA.
Voltage output
Move the cursor to the "Voltage output" menu item
using the
or
keys. Confirm your selection using
the key.
Use the
and keys to set the operating range of
the current output. Settings options: 0 V … 5 V or
0 V … 10 V.
Confirm your selection using the key. You will automatically return to the main menu.
Operating range 0…5 V:
Distance measurement:
Content measurement:
0 V
12 cm
0 V
0% Tank content
5 V
250 cm
5 V
100% Tank content
If "No echo received“: 0 V.
Operating range 0…10 V:
Distance measurement:
Content measurement:
0 V
12 cm
0 V
0% Tank content
10 V
250 cm
10 V
100% Tank content
If "No echo received“: 0 V.
Operating range of
current output:
0 mA .. 20 mA ™
Modbus settings
>
Current output
>
Voltage output
> ™
Relay settings
>
Operating range of
Voltage output:
0 V .. 10 V ™