1000356V01 DOC - ECOLOG-PRO RBR and 1NTR - Operation Manual_EN | Page 6 / 13
Operating altitude
This device must not be used at altitudes higher than 2000 meters above sea
level.
IP protection class
IP20
ID number
The ID number printed on the front of the module is the module's default host
name.
Communication interval
1 to 60 minutes, depending on configuration of sensors
Radio connection to measurement
modules
Available in two versions with 868 MHz (Europe) or 915 MHz (US)
Only the supplied antenna may be used.
Left LED
Red: Module is active, but is not connected to the network
Yellow: Module is active and is connected to the network
Green: Module is active and is connected to the monitoring software
Blue: Module is active and communicating
Right LED
Green: Module is connected to one or several measurement modules
Blue: Communication with measurement module taking place
Factory default
Press and hold the Factory Reset Button at the bottom for more than 3 seconds to
reset the module's factory settings:
-
Left LED shows three short flashes in green.
-
All measurement modules are disconnected.
-
Communication module is disconnected from the network.
-
Network settings are reset to default settings: module is configured for
dynamic IP addresses (DHCP client); host name equals ID number
Manual configuration of network
settings
Use the online tool available at
https://web.elpro.cloud/bridge
to manually configure specific network settings.
How to connect:
1.
Plug in your ECOLOG-PRO RBR using the power supply unit provided with your device.
1a. Once power is connected, the ECOLOG-PRO RBR's left LED will turn red.
2.
Connect the ECOLOG-PRO RBR to the Internet by connecting it to your modem, router or local
network.
2a. Once the connection to the network is made, the left LED will turn yellow, indicating that it is
connected to the network
2b. The left LED will ultimately turn green, indicating that the module is connected to ELPRO
Cloud.