Measurement Module ECOLOG-PRO 1NTR 

 

 

Functional characteristics: 

ECOLOG-PRO 1NTR is a measurement module with an internal temperature sensor. Communication 
from this module with the ECOLOG-PRO RBR is via radio signal. 
 

Power supply 

2 AAA Alkaline or LiFe (lithium-iron) batteries 
 
SAFETY INSTRUCTIONS! 
Improper use or misuse of batteries may result in leakage of battery fluid or, in 
the worst case, explosion and/or fire. Please note: 

-

 

Always choose the correct type of battery that best suits your intended 
use (see 'Measurement range').  

-

 

So that battery status can be determined correctly, always insert new, 
unused batteries when changing batteries.  

-

 

Never insert different types of batteries into the device at the same 
time. 

-

 

All batteries should be replaced at the same time. Mixing old and new 
batteries may cause the batteries to leak and damage the product. 

-

 

Make sure that the battery contacts and mating contacts in the device 
are clean before inserting the batteries. 

-

 

Make sure that the batteries are inserted properly i.e. observe polarity 
(+ and -). 

-

 

Remove the batteries if you are not going to use the product for a 
longer period of time to avoid damage from leakage. Leaking or 
damaged batteries can cause acid burns on skin contact. When handling 
damaged batteries, you should wear protective gloves. 

-

 

Remove expired batteries immediately from the device. 

-

 

Do not disassemble batteries, do not short-circuit them and do not 
throw them into fire. Never try to charge non-rechargeable batteries. 
There is a danger of explosion! 

-

 

Keep batteries out of reach of children. Do not leave batteries lying 
around as they may be swallowed by children or pets. 

Battery life 

14 months (provided that the module's radio connection is stable and that the 
defined temperature range (see below) is maintained) 

Measurement range 

 

With alkaline batteries:  -10°C.. +50°C 
With LiFe batteries: 

-35 °C..+50 °C 

Measurement resolution 

0.1 °C 

Measurement accuracy  

±1.0 °C in the range between -35.0 °C and -10.1 °C 
±0.5 °C in the range between -10.0 °C and -0.1 °C 
±0.4 °C in the range between 0.0 °C and +25.0 °C 
±0.8 °C in the range b25.1 °C and +50.0 °C 

Содержание ECOLOG-PRO 1NTR

Страница 1: ...PM PHIOSL 1000356V01 DOC ECOLOG PRO RBR and 1NTR Operation Manual_EN Page 1 13 ECOLOG PRO Radio Modules ECOLOG PRO RBR ECOLOG PRO 1NTR Operation Manual...

Страница 2: ...DOC ECOLOG PRO RBR and 1NTR Operation Manual_EN Page 2 13 Contents 1 System Overview 4 2 Communication Module ECOLOG PRO RBR 5 3 Measurement Module ECOLOG PRO 1NTR 7 4 Disposal 10 5 Declaration of Con...

Страница 3: ...owing internet address https www elpro com fileadmin Docs Quality_Documents ECOLOG PRO_RBR_DOC Conformity_1000278V01 pdf Above mentioned products operate in the 863 870MHz band with a maximum radiated...

Страница 4: ...lues are transmitted to a monitoring software ELPRO Cloud which stores and analyzes the data provides alerts if alarm limits are violated and generates reports The system provides superior visibility...

Страница 5: ...6 VDC 5 Output current 1200 mA max Country specific plugs Europe United Kingdom USA Size 55 x 31 x 41 mm Length of connector cable 1830 mm The module may only be operated with the supplied power supp...

Страница 6: ...g place Factory default Press and hold the Factory Reset Button at the bottom for more than 3 seconds to reset the module s factory settings Left LED shows three short flashes in green All measurement...

Страница 7: ...the device are clean before inserting the batteries Make sure that the batteries are inserted properly i e observe polarity and Remove the batteries if you are not going to use the product for a long...

Страница 8: ...to identify the module in the ELPRO Cloud software Radio connection to communication module Available in two versions 868 MHz Europe or 915 MHz US Left and middle LED Link Red flashing light on left L...

Страница 9: ...surement module for more than 3 seconds The left LED will blink red 3 times 4a Wait until the left and middle LEDs start to blink synchronously once per second 4b A successful connection will be indic...

Страница 10: ...ately from the product b Batteries You are legally obliged to dispose of all used batteries according to applicable laws disposal via household waste is prohibited Batteries categorized as containing...

Страница 11: ...1000356V01 DOC ECOLOG PRO RBR and 1NTR Operation Manual_EN Page 11 13 5 Declaration of Conformity...

Страница 12: ...1000356V01 DOC ECOLOG PRO RBR and 1NTR Operation Manual_EN Page 12 13...

Страница 13: ...1000356V01 DOC ECOLOG PRO RBR and 1NTR Operation Manual_EN Page 13 13 ELPRO BUCHS AG Lang ulistrasse 45 9470 Buchs SWITZERLAND E Mail swiss elpro com For local agencies see www elpro com...

Отзывы: