12. TEMPERATURE MEASUREMENT
The temperature measurement is made in the following way:
-
connect the temperature probe to the Chinch connector;
-
by pressing the
button switch the meter on;
-
choose the temperature measuring mode using the
button;
-
put the temperature probe to the measured solution;
-
wait till the value stabilises and read the result.
The meter co-operates with platinum resistor sensor PT-1000 and the final
accuracy of the temperature measurement depends on its class.
Pic. 14
NOTICE:
break in the circuit or disconnecting of the temperature probe is
signalised by disappearing of the . In such case the meter shows the
temperature value introduced by the user for manual temperature
compensation. Blinking -50°C value during measurement in positive
temperatures informs about short circuit in the temperature probe.
Содержание CC- 411
Страница 1: ...WATERPROOF CONDUCTIVITY METER CC 411 USER S MANUAL...
Страница 2: ......
Страница 3: ...USER S MANUAL WATERPROOF CONDUCTIVITY METER CC 411 Before use please read the instruction carefully...
Страница 4: ......
Страница 6: ......
Страница 7: ...1 I INTRODUCTION...
Страница 8: ...2 User s manual for CC 411 conductivity meter v5 30...
Страница 14: ...8 User s manual for CC 411 conductivity meter v5 30 Pic 2...
Страница 18: ...12 User s manual for CC 411 conductivity meter v5 30...
Страница 19: ...13 II CONDUCTIVITY MEASUREMENT...
Страница 20: ...14 User s manual for CC 411 conductivity meter v5 30...
Страница 34: ...28 User s manual for CC 411 conductivity meter v5 30...
Страница 35: ...29 III TEMPERATURE MEASUREMENT...
Страница 36: ...30 User s manual for CC 411 conductivity meter v5 30...
Страница 38: ...32 User s manual for CC 411 conductivity meter v5 30...
Страница 39: ...33 IV OTHER...
Страница 40: ...34 User s manual for CC 411 conductivity meter v5 30...
Страница 45: ...39...
Страница 46: ...40 User s manual for CC 411 conductivity meter v5 30...
Страница 48: ......
Страница 49: ......
Страница 50: ......
Страница 51: ......