8.  CHOICE AND MAINTENANCE OF THE CONDUCTIVITY CELL

8.1. Choice of the conductivity cell

Conductivity 

in 

CC-411

 

meter 

can 

be 

measured 

in 

range

0 – 100 mS/cm. The meter co-operates with conductivity cells with K constant
=  0.050 ÷ 1.500 cm-1  and  equipped  with  BNC-50  connector.  Depending  on
the required measuring range it is necessary to choose appropriate cell with K
constant which enables obtaining accurate results. Beyond the range the cell
looses its linearity and the results have greater error. Accurate measurements
in the whole range are possible with use of 2 different conductivity cells.
In range 1 – 100 mS/cm the standard conductivity cell with K constant =1 cm-

1 should be used.
For measurements of ultra pure water with conductivity below 1 

µ

S/cm a cell

with K constant = 0.1 cm-1 should be used.
Depending  on  the  expected  measuring  range,  choose  the  appropriate  cell
using the chart below (Pic.6).

Pic. 6. Dependence between the measuring range and of applied conductivity
cells.

Содержание CC- 411

Страница 1: ...WATERPROOF CONDUCTIVITY METER CC 411 USER S MANUAL...

Страница 2: ......

Страница 3: ...USER S MANUAL WATERPROOF CONDUCTIVITY METER CC 411 Before use please read the instruction carefully...

Страница 4: ......

Страница 5: ...9 Calibration of the conductivity cell 18 9 1 Calibration without the standard solution 18 9 2 Calibration with use of standard solution 19 10 Conductivity measurement 22 10 1 Conductivity measurement...

Страница 6: ......

Страница 7: ...1 I INTRODUCTION...

Страница 8: ...2 User s manual for CC 411 conductivity meter v5 30...

Страница 9: ...res obtaining high measuring parameters We want to call your attention to the fact that this equipment has much shorter working life than the meter In case of conductivity measurements it is important...

Страница 10: ...battery The meter is equipped with large custom LCD display which displays the conductivity or the temperature value Waterproof housing makes working in difficult conditions possible Minimised size a...

Страница 11: ...asurements in the laboratory Waterproof housing enables work in difficult weather conditions or in humid environment The CC 411 conductivity meter is used in food chemical pharmaceutical and power ind...

Страница 12: ...tton Symbols of units are displayed next to the result In case of disconnecting the temperature probe the meter switches to the manual temperature compensation mode Next to the result there will be di...

Страница 13: ...g and holding returns to the measurement mode holding this button enters the calibration mode CAL symbol displayed buttons used for entering the parameters In the upper wall of the meter there are inp...

Страница 14: ...8 User s manual for CC 411 conductivity meter v5 30 Pic 2...

Страница 15: ...ettings and requires service repair If after 1 5 s all symbols are continuously displayed it informs that the calibration parameters of the cell have been lost After pressing the button the meter adop...

Страница 16: ...nductivity at S cm or mS cm converted to 25o C salinity in g l converted to NaCl accordinfg to actual slope salinity in g l converted to TDS with use of the WTDS coefficient conductivity of honey at S...

Страница 17: ...re measured by the probe Disconnecting the temperature probe switches the meter to the manual temperature compensation mode the symbol disappears and the value of the temperature entered by the user w...

Страница 18: ...12 User s manual for CC 411 conductivity meter v5 30...

Страница 19: ...13 II CONDUCTIVITY MEASUREMENT...

Страница 20: ...14 User s manual for CC 411 conductivity meter v5 30...

Страница 21: ...the K constant introduced to the meter s memory and displays the result in units of conductivity S cm or mS cm Shortened symbol of the unit is displayed next to the result S or mS Conductivity is aff...

Страница 22: ...constant which enables obtaining accurate results Beyond the range the cell looses its linearity and the results have greater error Accurate measurements in the whole range are possible with use of 2...

Страница 23: ...ecommend universal liquid for cleaning cells This is a mixture of equal parts of isopropyl alcohol ethyl ether and hydrocholic acid diluted with water in 1 1 ratio Breaking of the measuring cell hinde...

Страница 24: ...ly determined by the cell manufacturer and using this value will be the most reliable In case of calibration performed by the user fresh accurately prepared standard solution is required Additionally...

Страница 25: ...measurement It is required to use standard solutions of a high quality 9 2 1 Entering the standard solution value To enter the standard solurion value choose the measurement unit according to the poin...

Страница 26: ...lled with the measured solution there shouldn t be any air bubbles and the electrodes should be evenly moistened measure the temperature of the solution and bring it to 25 oC in the conductivity measu...

Страница 27: ...y measuring mode press and hold the button till displaying the CAL symbol on the screen Pic 10 after the value stabilises the P1 symbol calibration point will be displayed Press the button Blinking of...

Страница 28: ...measuring mode according to the subchapter 6 1 if the conductivity cell has not been calibrated earlier calibrate it according to the chapter 9 place both probes in the measured solution the conductiv...

Страница 29: ...both probes in the measured solution the conductivity cell cannot touch the walls and the bottom The measuring cell should be filled with the sample solution and should not include any air bubbles the...

Страница 30: ...conductivity measurement unit according to the subchapter 6 1 if the conductivity cell has not been calibrated earlier calibrate it according to the chapter 9 measure the temperature of the solution...

Страница 31: ...of honey Table 1 Honey Conductivity x 10 4 S cm The meter s indication Fir honeydew 10 1mS cm Leaf honeydew 9 900 S cm Honeydew nectar 8 800 S cm Chestnut honey 10 1mS cm Linden honey 6 8 7 8 680 780...

Страница 32: ...by this coefficient and the result of salinity is received in g l e g if the conductivity value is 2 mS cm the salinity is 1g l In practice the dependence between conductivity and salinity isn t line...

Страница 33: ...ol on turn with the symbol displays 11 2 Salinity measurement with conversion to TDS The meter enables simplified way of determining the Total Dissolved Solids TDS content For calculations the constan...

Страница 34: ...28 User s manual for CC 411 conductivity meter v5 30...

Страница 35: ...29 III TEMPERATURE MEASUREMENT...

Страница 36: ...30 User s manual for CC 411 conductivity meter v5 30...

Страница 37: ...e stabilises and read the result The meter co operates with platinum resistor sensor PT 1000 and the final accuracy of the temperature measurement depends on its class Pic 14 NOTICE break in the circu...

Страница 38: ...32 User s manual for CC 411 conductivity meter v5 30...

Страница 39: ...33 IV OTHER...

Страница 40: ...34 User s manual for CC 411 conductivity meter v5 30...

Страница 41: ...ad of the display test the screen as in the picture below will appear Pic 15 In the upper row the software version will appear and in the lower row a type of internal power supply to which the meter h...

Страница 42: ...t the battery should be changed In order to do so it is necessary to undo two screws in the lower wall of the meter pull out the whole wall and replace the battery The next thing is to put the new bat...

Страница 43: ...es of this constant Temperature compensation manual automatic Compensation range 5 0 70 0 oC K constant range 0 05 1 500 cm 1 coefficient constant 2 oC WTDS coefficient constant 0 5 oC Cell calibratio...

Страница 44: ...SPLAY Custom LCD 55 x 45 mm DIMENSIONS 149 x 82 x 22 mm WEIGHT 220 g 16 EQUIPMENT Standard set 1 Conductivity cell 2 Pt 1000B temperature probe standard 3 9V battery 4 Plastic case for the meter cell...

Страница 45: ...39...

Страница 46: ...40 User s manual for CC 411 conductivity meter v5 30...

Страница 47: ...ormity with the users manual using wrong power adapter mechanical damages and damages caused by repairs made by unauthorised persons The conductivity cell and temperature probe have 12 month manufactu...

Страница 48: ......

Страница 49: ......

Страница 50: ......

Страница 51: ......

Страница 52: ...Sp j 41 814 Zabrze Witosa 10 POLAND tel 48 32 2738106 fax 48 32 2738114 www elmetron com e mail info elmetron com pl...

Отзывы: