- 23 -
10.2. Conductivity
measurement
with
automatic
temperature
compensation
In case of measurement with automatic temperature compensation:
-
connect the conductivity cell and the temperature probe to the connectors
F1
and
t
respectively (pic.2);
-
turn the meter on with the
button;
-
choose the conductivity measurement unit according to the subchapter 6.1;
-
if the conductivity cell has not been calibrated earlier, calibrate it according
to the chapter 9;
-
place both probes in the measured solution, the conductivity cell cannot
touch the walls and the bottom. The measuring cell should be filled with the
sample solution and should not include any air bubbles, the electrode’s
surface should be evenly moistened*;
-
wait till the result stabilises and read it (pic.12).
Pic. 12
Note:
in case of exceeding the range of temperature compensation the
conductivity reading and the symbol start blinking
* - air bubbles may be removed by moving the immersed cell. In order to make moistening of the electrodes
easier, it is recommended to immerse the cell in distilled water with washing-up liquid, and than wash it in
distilled water.
Содержание CC- 411
Страница 1: ...WATERPROOF CONDUCTIVITY METER CC 411 USER S MANUAL...
Страница 2: ......
Страница 3: ...USER S MANUAL WATERPROOF CONDUCTIVITY METER CC 411 Before use please read the instruction carefully...
Страница 4: ......
Страница 6: ......
Страница 7: ...1 I INTRODUCTION...
Страница 8: ...2 User s manual for CC 411 conductivity meter v5 30...
Страница 14: ...8 User s manual for CC 411 conductivity meter v5 30 Pic 2...
Страница 18: ...12 User s manual for CC 411 conductivity meter v5 30...
Страница 19: ...13 II CONDUCTIVITY MEASUREMENT...
Страница 20: ...14 User s manual for CC 411 conductivity meter v5 30...
Страница 34: ...28 User s manual for CC 411 conductivity meter v5 30...
Страница 35: ...29 III TEMPERATURE MEASUREMENT...
Страница 36: ...30 User s manual for CC 411 conductivity meter v5 30...
Страница 38: ...32 User s manual for CC 411 conductivity meter v5 30...
Страница 39: ...33 IV OTHER...
Страница 40: ...34 User s manual for CC 411 conductivity meter v5 30...
Страница 45: ...39...
Страница 46: ...40 User s manual for CC 411 conductivity meter v5 30...
Страница 48: ......
Страница 49: ......
Страница 50: ......
Страница 51: ......