5. SETTINGS
5.1 DEFAULT SETTINGS
When delivered, 1005RE have the following default settings.
Output
0 - 20 mA
Alarm delay
300 sec.
Shutdown at alarm
Off
Offset 5%
Slope
100% ( 20 °)
Limit 1 (L1)( low )
100%
Limit 2 (L2)( high )
100%
Limit 3 (X1)( low )
100%
Limit 4 (X2)( high )
100%
Regulation On/Off
Xp-band 100.0%
Frequency
180 pulses/ min
PWM 100
sec
I-Time 2000
sec
5.2 PARAMETER SETTINGS
All settings has to be confirmed with
=
to be saved. To leave the setting mode without saving the
settings: press
M
.
Before starting up the controller the following parameters has to be set:
Turn the power supply on.
5.2.1 Relay configuration
When delivered the instrument is pre-set for On/Off regulation. Limit value 1 is set to 100%, min.value
(
LO
), limit value 2 is set to 100%, min.value (
LO
), limit value 3 is set to 100%, max.value (
HI
) and
limit value 4 is set to pH = 100%, max.value (
HI
).
5.2.2a) Setting of limit values
NOTE! Relay 3 and 4 (L1 and L2) provides only ON/OFF-function.
1. Interrupt the measuring by pressing
+
and
-
at the same time until the measuring is
interrupted and the diode at
TEMP
is activated.
2. Step
to
L1
with
M
.
3. On
display:
100.0
. If another limit value is desired, use
+
and
-
to set a new value.
Confirm with
=
. On display:
-LO-
.
4. Choose if limit value 1 is a min
( -LO- )
or max-function
( -HI- )
,
use
+
. Press
=
to
confirm.
5. The diode at
L2
is activated. On display:
100.0
. If another limit value is desired, use
+
and
-
to set a new value. Confirm with
=
. On display:
-LO-
.