3. FUNCTIONS
3.1 COMMON FUNCTIONS
All settings are made through the function keys on the front panel. The keys
+
and
–
are accelerating
in three steps when they are kept pressed.
Chosen function is indicated by a green diode in the menu. At alarm, a red diode is activated at the
same time as the diode at the actual function activates. When pressing
M
you step down in the menu.
When calibration or setting mode is entered, the output signal stays "frozen" at the latest value, the
relays opens up and the regulation is paused, until
=
is pressed.
3.2
3.2.1 MEASURE
When power supply is connected the measuring mode starts.
3.2.2 CAL. 1
Offset calibration. Pre-set value : 5%.
3.2.3 EXIT TO MEASURE
Returns to measuring mode.
3.2.4 L1
Settings of limit value 1 (on/off-output) and alarm limit 1. The value can be set between 0,5 and 99,5%
3.2.5 L2
Setting of limit value 2 (on/off-output) and alarm limit 2. The value can be set between 0,5 and 99,5%.
3.2.6 X1
Setting of limit value 3. The value can be set between pH 0,5 and 99,5%.
3.2.7 X2
Setting of limit value 4. The value can be set between pH 0,5 and 99,5%.
3.2.8 PWM
Pulse length, between 5 – 100 seconds.
X1
X2
PWM
FRK
P-BAND %
ALARM
OUTPUT
CAL1 Offset
EXIT TO MEASURE
L1
L2
I-TIME sec.