16 Maintenance
Elma Schmidbauer GmbH
30 / 32
Elmasonic Select
Display
Fault type
Possible remedies
Er 40–44
Power supply faulty.
Connect the unit to a socket of a different power
circuit.
Er 47
Controls overheated.
Switch off unit and allow to cool. Elmasonic Select
120–900: check that the blower is clean and
working correctly. Remove obstructions from
below the unit.
Table 3:
Error messages
Contact the service centre or the manufacturer in the following cases.
• The measures described here do not remedy the faults.
• The faults recur or recur repeatedly, even after performing the following measures:
– Disconnect the unit from the mains power, then reconnect and restart the unit after approx. one
minute.
– The unit has been reset to factory settings.
• The error number displayed is not listed here: make a note of the error number; the unit must be
repaired at the factory.
IMPORTA
NT
Empty the unit completely, and clean and disinfect it thoroughly before sending it for
repair.
See also
2
16
Maintenance
CAUTION
Faulty network cable
Electric shock or damage to the unit
Ø
Check the network cable regularly for damage.
IMPORTA
NT
Keep the unit and tank clean and dry to prolong its service life. Regularly remove all dirt
residues and accumulated dust from the blower.
16.1
Replace the network cable
ü
The network cable or plug contacts are visibly damaged.
◦ Replace damaged network cables immediately.
NOTICE! Removable network cables may not be replaced with insufficiently measured
network cables!
è
The network cable has been replaced.