Elma 117 - 118
Page 57
Capacitance
Range
ACCURACY
1)
100.0NF
2)
, 1000NF,
10.00
μ
F, 100.0
μ
F,
2000
μ
F
3.5%+5D
3)
1)
ACCURACIES WITH FILM CAPACITOR OR BETTER
2)
ACCURACY BELOW 50NF IS NOT SPECIFIED
3)
Specified with battery voltage above 2.8V (approximately half full battery). Accuracy decreases
gradually to 12% at low battery warning voltage of approximately 2.4V
Non-Contact EF-Detection
Typical Voltage
Bar Graph
Indication
15V to 85V
-
40V to 130V
- -
60V to 210V
- - -
90V to 300V
- - - -
above 120V
- - - - -
Indication: Bar graph segments & audible beep tones proportional to the field strength
Detection Frequency: 50/60Hz
Detection Antenna: Top side of the stationary jaw
Probe-Contact EF-Detection: For more precise indication of live wires, use the Red (+) probe for
direct contact measurements
ACA CURRENT (CLAMP-ON)
RANGE ACCURACY
1) 2) 3)
50HZ / 60HZ
400.0A, 2000A
1.5% + 5D
True RMS model BM118 Crest Factor:
< 2.0 : 1 at full scale & < 4.0 : 1 at half scale
1)
Add 8d to specified accuracy while reading is below 10% of range
2)
Induced error from adjacent current-carrying conductor: < 0.06A/A
3)
Specified accuracy is for measurements made at the jaw center. When the conductor is not
positioned at the jaw center, position errors introduced are:
ADD 1% TO SPECIFIED ACCURACY FOR MEASUREMENTS MADE WITHIN JAW MARKING
LINES (AWAY FROM JAW OPENING)
Add 4% to specified accuracy for measurements made BEYOND jaw marking lines (toward jaws
opening)
Содержание Auto-Check 117
Страница 12: ...Elma 117 118 Sida 11 BRUKSANVISNING Elma 117 118 T ngmultimeter med AutoCheckTM...
Страница 24: ...Elma 117 118 Sivu 23 K ytt ohje Elma 117 118 Digitaalinen pihtimittari Elma 117 El nro 63 98 910 152...
Страница 34: ...Elma 117 118 Seite 33 Bedienungs anleitung Elma 117 118 Multifunktions Zangen Amp remeter...
Страница 44: ...Elma 117 118 Page 43 USER S MANUAL Elma 117 118 Auto CheckTM Clamp on Multimeter Series...
Страница 60: ......