1. Safety Warnings

  This  instrument  has  been  designed,  manufactured  and 

tested according to the following standards.

  IEC 61010-1 Measurement CAT

 300V Pollution degree 2

  IEC 61010-031

  IEC 61326

   This  instruction  manual  contains  warnings  and  safety 

rules  which  must  be  observed  by  the  user  to  ensure 

safe  operation  of  the  instrument  and  retain  it  in  safe 

condition.  Therefore,  read  through  these  operating 

instructions before using the instrument.

#

 

WARNING

●  Read through and understand instructions contained 

in this manual before starting using the instrument.

●  Save  and  keep  the  manual  handy  to  enable  quick 

reference whenever necessary.

●  This  instrument  should  only  be  used  by  suitably 

trained  person  and  be  sure  to  follow  measurement 

procedures  described  in  the  manual.  Kyoritsu 

assumes  no  responsibility  for  damage  and  injury 

caused by misuse or not following instructions in the 

manual.

●  Be  sure  to  understand  and  follow  all  safety 

instructions contained in the manual.

Be sure to observe the above instructions.

 Failure  to  follow  the  above  instructions  may  cause 

injury, instrument damage and/or damage to equipment 

under test.

The  symbol 

#

 

indicated  on  the  instrument  means  that 

the user must refer to related parts in the manual for safe 

operation of the instrument. 

Be sure to carefully read the instructions following 

#

 

each 

symbol in the manual.

#

 

 

DANGER

  is reserved for conditions and actions that 

are likely to cause serious or fatal injury.

#

 

 

WARNING

    is  reserved  for  conditions  and  actions 

that can cause serious or fatal injury.

#

 

 

CAUTION

  is reserved for conditions and actions that 

can cause injury or instrument damage.

#

 

DANGER

●  Never make measurement on the circuit in which electrical 

potential to ground over 300V AC/DC exists.

●  Do not attempt to make measurement in the presence of 

flammable gasses. Otherwise, the use of the instrument 

may cause sparking, which can lead to an explosion.

●  Be sure to keep your fingers behind the Finger barrier 

part of test lead.

●  Never  attempt  to  use  the  instrument  if  its  surface  or 

your hand is wet.

●  Do  not  open  the  instrument  case  when  making 

measurement.

#

 

WARNING

●  Never  attempt  to  make  any  measurement  if  any 

abnormal conditions are noted, such as broken case, 

cracked test leads and exposed metal parts.

●  Do  not  turn  the  function  selector  switch  with  test 

leads connected to the instrument.

●  Do  not  install  substitute  parts  or  make  any 

modification to the instrument. 

●  Do not try to replace the batteries if the surface of the 

instrument is wet.

●  Make  sure  to  disconnect  test  leads  from  the  device 

under  test  when  opening  the  case  for  battery 

replacement.

● Use of our Protective Cap offers different lengths of the 

exposed metal part suitable for the test environments. 

  Please attach the Cap onto the metal part under CAT.

III or higher test  environments.

#

 

CAUTION

●  Always make sure to check Function switch is setting 

to the appropriate range before starting measurement.

●  Do not expose the instrument to the direct sun, high 

temperature and humidity or dewfall.

●  When  the  instrument  will  note  be  in  use  for  a  long 

period, place it in storage after removing the batteries.

●  Use  a  cloth  dipped  in  water  or  neutral  detergent  for 

cleaning the instrument. Do not use abrasives or solvents.

●The Cap should be firmly attached to the Probes.
 Please  refer  to  following  explanation  of  the  symbols 

marked on the instrument or in the manual.

●  Symbols
 

 

Ground 

 

AC   

DC

 

  :AC and DC 

Ω :

Resistance

 

 

Diode 

  :Buzzer

 

 :Capacitor 

Hz :

Frequency

 

 

Double or reinforced insulation

● 

Measurement

  CAT

: 

Primary electrical circuit of equipment with 

power cord for connection to outlet.

  CAT

: 

P r i m a r y  e l e c t r i c a l  c i r c u i t  o f  t h e 

equipment,  which  is  supplied  power  from 

a  distribution  board,  and  cable  run  from  a 

distribution board to an outlet.

2. Features

  This is a Digital Multi Meter providing most portability 

by means of stowing the instrument body together with 

probe in the notebook-size cover.

 ● Designed to international safety standards. 

   IEC 61010-1 Measurement CAT.

 300V 

   Pollution degree 2  

   IEC 61010-031 (probe assemblies)

 ●  REL  function  to  check  the  difference  of  measured 

values

 ●  Auto power off function to save battery consumption

 ● Data hold function

 ● Diode and Continuity check function

 ● Auto-ranging function

 ● Frequency measurement function

 ● DUTY measurement function

    (Express  Pulse  width  /  Pulse  period  as  a 

percentage)

3. Specification

 ● Measuring ranges and accuracy

   (23±5℃ , under the 45%〜75%RH)

DCV Function(5 Autoranging) :Input impedance approx. 10MΩ

Range

Measuring range

Accuracy

400mV

0〜600V

 ±0.8%rdg±5dgt

4V

40V

400V

600V

 ±1.0%rdg±5dgt

ACV Function(4 Autoranging): Input impedance approx. 10MΩ

Range

Measuring range

Accuracy

4V

0〜600V

 ±1.3%rdg±5dgt (50 / 60Hz)

 ±1.7%rdg±5dgt (〜400Hz)

40V

400V

 ±1.6%rdg±5dgt (50 / 60Hz)

 ±2.0%rdg±5dgt (〜400Hz)

600V

Resistance Function(6 Autoranging)

Range

Measuring range

Accuracy

400Ω

0〜40MΩ

 ±1.0%rdg±5dgt

4kΩ

40kΩ

400kΩ

4MΩ

40MΩ

 ±2.5%rdg±5dgt

Diode check/Continuity check Function

Function

Measuring range

Diode check

Test current approx. 0.4mA

Continuity check

Buzzer beeps below about 120Ω

Capacity Function (6 Autoranging)

Range

Measuring range

Accuracy

4nF

〜200uF

 ±5.5%rdg±10dgt 

40nF

 ±3.5%rdg±10dgt 

400nF

 ±3.5%rdg±5dgt 

4uF

40uF

200uF

 ±4.5%rdg±5dgt 

Frequency (4 Autoranging)/ DUTY Function

Range

Measuring range

Accuracy

10Hz

〜10kHz

Input sensitivity: 

more than 1.5V(RMS)

 ±0.1%rdg±5dgt

100Hz
1000Hz
10kHz
DUTY

0.1〜99.9%

(Pulse width / Pulse period)  ±2.5%rdg±5dgt

Note:

* At Voltage function, the Auto-ranging function is released 

by pressing the SELECT key. To measure a voltage again, 

turn the Function switch to the "OFF" position once. Then 

set it to the Voltage function again.

● Standards :

  IEC61010-1  Measurement CAT

 300V, pollution degree 2

  IEC61010-031

  capped condition for CAT.III 300V

  uncapped condition for CAT.II 300V

  IEC61326

● Method of operation : 

△∑method

● Indication: 

  LCD maximum value 3999 (ACV, DCV, Ω, F)

  units, symbols

● Over range display : 

  "OL" symbol is displayed on the LCD.

   In  case  that  the  value  is  beyond  effective  measuring 

range  at  the  position  of  Ω  function  range  and  manual 

range.

● Auto-ranging : 

   Range  shifts  to  upper  range  when  indicated  value  is 

more  than  3999.  Range  shifts  to  lower  range  when 

indicated value is less than 360.

● Sampling rate :

 approx. 400ms

● Operating Environmental conditions

   −indoor use

   −altitude up to 2000m

● Temperature & Humidity range(guaranteed accuracy):

  23℃±5℃   Relative humidity: less than 75%

● Operating Temperature & Humidity range:

  0℃〜+40℃  Relative humidity: less than 80%

● Storage Temperature & Humidity range:

  -20℃〜+60℃  Relative humidity: less than 70%

● Insulation Resistance: 

   It  should  be  more  than  10MΩ/DC1000V  between 

electrical circuit and enclosures.

● Withstand Voltage : 

   

   It  should  be  more  than  AC3700V/  for  one  minute 

between electrical circuit and enclosures.

● Overload Protection :

  Voltage function : 

720V(RMS.) 10seconds

  Resistance function : 250V(RMS.) 10seconds

  Diode / Continuity :  250V(RMS.) 10seconds

  Capacity function :  250V(RMS.) 10seconds

  Frequency function :  250V(RMS.) 10seconds

● Dimensions : 

 approx. 107(L) x 54(W) x 10(D) mm

● Weight :

  approx. 70g(including batteries)

● Power source :

  Two LR44(SR44)1.5V or equivalent

● Accessories :

  Two LR44 (SR44) batteries

 

       Portable holder

 

       Instruction manual

#

 

CAUTION

The  voltage  shown  above  is  the  overload  protection 

(overvoltage protection) for the instrument.

Make sure not to exceed the value of voltage shown 

above.

4. Instrument Layout

① Function Selector Switch 

④DATA HOLD Key 

② Display   

 

⑤Test Leads

③ SELECT Key

5. Preparation

 5-1 Checking Battery Voltage

   Set  the  Function  Selector  Switch  to  other  positions 

except the OFF position.

   Battery  Voltage  is  enough  if  indication  is  clear  and  

BT symbol is not indicated in this bout.

   If  BT  symbol  is  indicated  or  no  indication  on  the 

display, follow to the Battery Replacement procedures 

shown in item9 in this document and replace with new 

batteries.

INSTRUCTION MANUAL

6. Measurements

 6-1VoltageMeasurement(DCV,ACV)

#

 

DANGER

●  To  avoid  the  danger  of  getting  electrical  shock, 

never  make  measurement  on  a  circuit  over  600V 

AC/DC. (electrical potential to ground 300V AC/DC)

●  Do  not  operate  Function  Selector  Switch  during 

measurement.

●  Do  not  make  measurement  when  opening  the 

instrument case.

 6-1-1  DC Voltage Measurement (DCV)

 (1) Set the Function Selector Switch to " " position.

      (Then,  "AUTO"," "and"mV"  symbols  are  indicated 

on the display.)

 (2)  Connect  the  black  test  lead  to  the  negative  side  of 

the circuit under the test and the red test lead to the 

positive side of the circuit, then the measured value 

is  indicated  on  the  display.  If  you  connect  the  test 

leads  the  other  way,  "-"  symbol  is  indicated  on  the 

display.

 6-1-2  AC Voltage Measurement (ACV)

 (1) Set the Function Selector Switch to " " position.
      (Then, "AUTO", "

", and "V" symbols are indicated 

on the display.)

 (2) Connect the test leads to the circuit under test.

    Measured value is indicated on the display.

Note)  Even  if  short-circuit  the  input  line  at  the  range  of 

AC4V, 2〜5dgt may remain indicated.

 6-2ResistanceMeasurement

#

 

DANGER

●  To  avoid  the  danger  of  getting  electrical  shock, 

never  make  measurement  of  the  circuit  in  which 

electric potential exists.

●  Do  not  make  measurement  when  opening  the 

instrument case.

 (1)  Set the Function Selector Switch to "Ω" position.

      (Then,  "AUTO"  and  "MΩ"  symbols  are  indicated 

on  the  display.)  Make  sure  that  the  "OL"  symbol 

is  indicated  on  the  display  at  this  bout,  then  short 

the test lead tips and check "0" is indicated on the 

display.

 (2)  Connect the test leads to both ends of the resistance 

under test.

     Measured value is indicated on the display.

Note)  Even if short the test lead tips, indicated value may 

not be "0". But this is because of the resistance of 

test leads and not a failure. 

 6-3ContinuityCheck/DiodeCheck

#

 

DANGER

●  To  avoid  the  danger  of  getting  electrical  shock, 

never  make  measurement  of  the  circuit  in  which 

electric potential exists.

●  Do  not  make  measurement  when  opening  the 

instrument case.

 6-3-1  Continuity Check

 (1)  Set the Function Selector Switch to "

/ " position.

      (Then,  " "  and  Ω  symbols  are  indicated  on  the 

display.)

 (2)  Make  sure  that  the  "OL"  symbol  is  indicated  on  the 

display at this bout, then short the test lead tips and 

check "0" is indicated on the display and check if the 

buzzer beeps.

 (3)   Connect the test leads to both ends of the resistance 

under test. 

      Measured  value  is  indicated  on  the  display.  The 

buzzer beeps below about 120Ω.

Note)  Even if short the test lead tips, indicated value may 

not be "0". But this is because of the resistance of 

test leads and not a failure.

 6-3-2  Diode Check

 (1) Set the Function Selector Switch to "

/ " position.

      (Then,  " "  and  "Ω"  symbols  are  indicated  on  the 

display.)

 (2)  Press the SELECT Key twice and set the instrument 

to Diode Check mode. (Then, "

" and "V" symbols 

are  indicated  on  the  display.)  Make  sure  that  the 

"OL"  symbol  is  indicated  on  the  display  at  this 

bout,  then  short  the  test  lead  tips  and  check  "0"  is 

indicated on the display.

 (3)  Connect  the  black  test  lead  to  the  cathode  side  of 

the Diode and the red test lead to the anode side of 

the Diode.

      Forward voltage of Diode is indicated on the display.

 (4)  Connect the black test lead to the anode side of the 

Diode  and  the  red  test  lead  to  the  cathode  side  of 

the Diode.Normally, "OL" symbol is indicated on the 

display.

Conclusion :  Diode  is  OK  if  the  instrument  complies  with 

above items(3) and (4).

Note)  Open-circuit voltage between measuring terminals is 

approx.1.5V.(measuring current approx. 0.4mA) 

 6-4CapacitanceMeasurement

#

 

DANGER

●  To  avoid  the  danger  of  getting  electrical  shock, 

never  make  measurement  of  the  circuit  in  which 

electric potential exists.

●  Do  not  make  measurement  when  opening  the 

instrument case.

●  Make sure to discharge the capacitor before making 

measurement.

 (1) Set the Function Selector Switch to " " position.
      (Then,  "AUTO"  and  "nF"  symbols  are  indicated  on 

the display.)

 (2) Press the SELECT key and "0" shall be indicated.

      (Then, "△" symbol is indicated on the display.)

 (3)  Connect the test leads to both ends of the resistance 

under test.

      Measured value is indicated on the display. 

      Measuring  unit  "nF"  /  "uF"  is  automatically  chosen 

and indicated due to the measured value.

Note)  It  may  take  some  time  according  to  the  measuring 

capacitance.

      Measuring  capacitance  <  4uF  Measuring  time  is 

about 2seconds

      Measuring  capacitance  <  40uF  Measuring  time  is 

about 7seconds

      Measuring  capacitance  <  100uF  Measuring  time  is 

about 15seconds

 6-5FrequencyMeasurement

#

 

DANGER

●  To  avoid  the  danger  of  getting  electrical  shocks, 

never  make  measurement  on  a  circuit  over  300V 

AC/DC. (electrical potential to ground 300V AC/DC)

●  Do  not  operate  Function  Selector  Switch  during 

measurement.

●  Do  not  make  measurement  when  opening  the 

instrument case.

  Frequency  can  be  measured  at  ACV  functions  by 

pressing  "Hz/DUTY"  Switch.  Concerning  with  the 

direction  for  use  of  "Hz/  DUTY"  Switch,  please 

reference to the item7-1 Hz/DUTY in this document.

Note)  The minimum input can be measured is approx. 1.5V. 

     To measure a frequency, measure the voltage on the 

electrical  circuit  in  advance.  Then  press  the  "Hz/

DUTY" Key to enter into frequency measurement.

     Readings  of  frequency  may  fluctuate  or  be 

influenced under noisy environment.

7. How to use Function Switches

 7-1SELECT,Hz/DUTYKey

  At each function, the actions of SELECT, Hz/DUTY key 

are different so please refer to following items and make 

active use of them. 

 ● ACV function 

   (Above two keys act as Hz/DUTY switch key.)

    Capable  of  selecting  the  Voltage,  Frequency  or 

DUTY measurement mode. 

    At  the  initial  condition,  Voltage  measurement  has 

been selected for the ACV function.

    By  pressing  "Hz/DUTY"  key,  measuring  mode 

changes.

   "Voltage"    →    "Frequency"    →    "DUTY"
 ●  *At  Voltage  function,  the  Auto-ranging  function  is 

released  by  pressing  the  SELECT  key.  To  measure  a 

voltage  again,  turn  the  Function  switch  to  the  "OFF" 

position once. Then set it to the Voltage function again. 

 ● DCV, Ω and Capacitance measurement 

   (Above two keys act as REL△ key.)  

   Indicate the difference between measured values.

    When  any  function  ("DCV",  "Ω"  and  "Capacitance") 

selected,  the  measured  value  can  be  stored  by 

pressing  SELECT  key  and  after  that,  the  difference 

between  stored  value  and  measuring  value  is 

indicated on the display

    (△mark is keep lighting on the display while a value 

is stored)

    The  stored  value  can  be  released  by  pressing 

"SELECT" key again.

   "Release"     →     "Memory"
 ● Continuity check /Diode check function

   (Change between Continuity check and Diode check)

    At the initial condition, "Continuity check" mode has 

been selected for the Continuity check / Diode check 

function.

    By  pressing  "SELECT"  key,  measuring  mode 

changes.

   "Continuity check"     →     "Diode check"

    The relative measurement is allowed in the following 

range.

  * Measuring  range  =  Full  scale  value  at  a  range  − 

initial value

 7-2DATAHOLDKey

  The measured value can be hold at all functions.

  By  pressing  "DATA  HOLD"  key,  "DH"  symbol  indicated 

on the display and the indicated value can be held. 

  By  pressing  "DATA  HOLD"  key  again,  "DH"  symbol 

disappears from the display and held data is released. 

8. Auto Power Off

  Auto power off function operates when about 15minutes 

passed  after  power  on  this  instrument.  When  Auto 

power off function operates and the instrument powered 

off,  the  power-off  statue  returns  to  normal  by  pressing 

any key. 

9. Battery Replacement

#

 

DANGER

●  Never  open  the  instrument  case  when  making 

measurement.

●  To avoid getting electrical shock, be sure to remove 

test  leads  from  the  instrument  when  opening  the 

instrument case in order to replace batteries.

 (1) Remove the Portable holder from the instrument.

 (2)  Loosen  one  screw  on  the  bottom  of  the  instrument 

and open the battery cover, then replace batteries.

   Battery : Two LR44(SR44)1.5V or equivalent

10. Maintenance

  Use  a  cloth  dipped  in  water  or  neutral  detergent  for 

cleaning  the  instrument.  Do  not  use  abrasives  or 

solvents.

DISTRIBUTOR

Kyoritsu  reserves the rights to change specifications 

or  Designs  described  in  this  manual  without  notice 

and without obligations.

CARD TYPE  AUTO RANGE  DIGITAL  MULTIMETER

KEW 1018