7.
Turn on light source, move X.Y axis . CCD align 4-6 amount position
8.
Adjust target size, adjust initial value to target size too
9.
Revise coordinate
10.
Red disk will show, left key to save system file
H1 valve calibration:M12_cleaning point setting
When to use : Renewing X.Y.Z axis hardware for example motor screw, cleaning position is wrong and
therefore needs reset
1.
Select M12_H1 cleaning point setting
2.
Move X.Y.Z axis, align needle with vacuum suck cleaning needle as below Pic. 1
3.
Click Vacuum clean pin
4.
Revise
5.
Red disk will show, click left key to save system file
1
3
4
5