PURELAB 7000 - US Operator Manual
ELGA
Page 22
PURELAB 7000 - US MANU40676 Version 02 - 02/16
6. SANITIZATION
6.1
Liquid sanitization
RO sanitization should be completed on a regular basis or if water
quality deteriorates.
WARNING! ALWAYS WEAR RUBBER GLOVES, APRON
AND FACEMASK. DO NOT BREATHE FUMES
OR ALLOW TO COME IN CONTACT WITH SKIN
OR
EYES.
ALWAYS
FOLLOW
RECOMMENDATIONS
FOUND
IN
THE
MANUFACTURERS MATERIAL SAFETY DATA
SHEET
AND
ANY
APPLICABLE
OSHA
STANDARDS FOR THE CHEMICAL BEING
USED.
WARNING! LABEL THE MACHINE WITH APPROPRIATE
WARNING SIGNS SUCH A
S “DO NOT USE/
CONTAINS STERILANT” (NOT PROVIDED)
6.1.1 RO sanitization
Minncare Cold Sterilant is a Peracetic and Hydrogen Peroxide based
solution.
Refer to the Minncare label for additional information and follow all
applicable directions for use on the manufacturer’s label in
conjunction with the following instructions.
1.
ISOLATE application and ensure the process is OFF.
2.
PRESENT GREEN sanitization key to reader.
Present the sanitization key