Page 21
5.6
Replacing E - cartridge (LC181)
The E - cartridge should be replaced:
When indicated by the alarm.
1. ENSURE process is OFF and ISOLATE power.
2. OPEN front doors and LOCATE E - cartridge as
shown.
3. DISCONNECT tubing to air pump and inlet and
outlet tubing from the cartridge.
4. REMOVE cartridge from clip and discard.
5. REMOVE new E - cartridge from packaging.
6. REMOVE sealing plugs from inlet and outlet
ports.
7. REMOVE caps from cartridge side adaptors.
8. LOCATE cartridge in clip.
9. RE-CONNECT tubing.
5.7 RO modules
The Reverse Osmosis modules should be replaced if
permeate water purity or flow rate is not adequate and
does not meet predicted or previous performance.
For information regarding the replacement of the
Reverse Osmosis module contact Customer Services.