PURELAB 7000 - US Operator Manual
ELGA
Page 12
PURELAB 7000 - US MANU40676 Version 02 - 02/16
Step 18 - Data output (screen 035)
The
PURELAB
7000
can transmit operational data to an
Xport storage device.
1.
PRESS
to change (
= ON)
OR
PRESS
✔
to proceed to Step 19.
2.
PRESS
✔
.
Step 19 - Data transmit (screen 036)
Setting the frequency of data transmittals to Xport.
1.
PRESS
to change transmit intervals 1, 5, 15,
30 minutes, 1 hour or 6 hours
OR
PRESS
✔
to proceed to Step 20.
2.
PRESS
✔
.
Note:
PRESS PRINT during normal operation and
current data will be transmitted if data output
(screen 035) is selected.
Step 20 - Programming of User Keys (screen 037)
During the life of the
PURELAB 7000
you may need to
delete or add User keys to prevent or allow access to user
choices. This feature is only available to the Master key
holder.
CAUTION!
Do not press reset unless all User Keys
are present.
1.
PRESS
to delete all User keys.
2.
PRESENT the new User key to the reader.
3.
PRESS
to load new User key identification.
4.
REPEAT 2 and 3 until all User keys are registered
(maximum of 6 users).
5.
PRESS
✔
to complete settings.
Data transmit screen
Programming of user keys screen
Data output screen