(1 Related to the nominal values, the accuracy defines the maximum deviation between an adjusted values and the true (actual) value.
Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to differ max. 80 mV, which
means it might be between 4.92 V and 5.08 V.
(2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz
(3 The display error adds to the error of the related actual value on the DC output
1000 / 1500 W
Model
PS 9500-06 T
PS 9040-60 T
PS 9080-60 T
AC Input
Voltage range without derating
90...264 V AC
150...264 V AC
150...264 V AC
Voltage range with derating
-
90...150 V AC
90...150 V AC
Connection
1ph,N,PE
1ph,N,PE
1ph,N,PE
Frequency
45-65 Hz
45-65 Hz
45-65 Hz
Fusing
T 16 A
T 16 A
T 16 A
Leak current
< 3.5 mA
< 3.5 mA
< 3.5 mA
Power factor
~ 0.99
~ 0.99
~ 0.99
DC Output
Max. output voltage U
Max
500 V
40 V
80 V
Max. output current I
Max
6 A
60 A
60 A
Max. output power P
Max
1000 W
1500 W
1500 W
Max. output power P
Max
with derating
-
1000 W
1000 W
Overvoltage protection range
0...550 V
0...44 V
0...88 V
Overcurrent protection range
0...6.6 A
0...66 A
0...66 A
Overpower protection range
0…1100 W
0…1650 W
0…1650 W
Output capacitance
130 μF
6120 μF
6120 μF
Temperature coefficient for set values Δ/K
Voltage / current: 100 ppm
Voltage regulation
Adjustment range
0...500 V
0...40 V
0...80 V
Accuracy
(1
(at 23 ± 5°C)
< 0.1% U
Max
< 0.1% U
Max
< 0.1% U
Max
Line regulation at ±10% ΔU
AC
< 0.02% U
Max
< 0.02% U
Max
< 0.02% U
Max
Load regulation at 0...100% load
< 0.05% U
Max
< 0.05% U
Max
< 0.05% U
Max
Rise time 0...100% (@full load)
Max. 30 ms
Max. 40 ms
Max. 40 ms
Settling time after load step
< 1.5 ms
< 1.5 ms
< 1.5 ms
Display: Resolution
See section
„1.9.5.4. Resolution of the displayed values“
Display: Accuracy
(3
≤ 0.2% U
Max
≤ 0.2% U
Max
≤ 0.2% U
Max
Ripple
(2
< 155 mV
PP
< 33 mV
RMS
< 25 mV
PP
< 4 mV
RMS
< 25 mV
PP
< 4 mV
RMS
Remote sensing compensation
Max. 5% U
Max
Max. 5% U
Max
Max. 5% U
Max
Output voltage fall time (at no load) after
switching DC output off
Down from 100% to
<60 V: less than 10 s
-
Down from 100% to <60
V: less than 10 s
Current regulation
Adjustment range
0...6 A
0...60 A
0...60 A
Accuracy
(1
(at 23 ± 5°C)
< 0.2% I
Max
< 0.2% I
Max
< 0.2% I
Max
Line regulation at ±10% ΔU
AC
< 0.05% I
Max
< 0.05% I
Max
< 0.05% I
Max
Load regulation at 0...100% ΔU
OUT
< 0.15% I
Max
< 0.15% I
Max
< 0.15% I
Max
Ripple
(2
< 8 mA
RMS
< 6 mA
RMS
< 6 mA
RMS
Display: Resolution
See section
„1.9.5.4. Resolution of the displayed values“
Display: Accuracy
(3
≤ 0.2% I
Max
≤ 0.2% I
Max
≤ 0.2% I
Max
Power regulation
Adjustment range
0…1000 W
0…1500 W
0…1500 W
Accuracy
(1
(at 23 ± 5°C)
< 1% P
Max
< 1% P
Max
< 1% P
Max
Line regulation at ±10% ΔU
AC
< 0.05% P
Max
< 0.05% P
Max
< 0.05% P
Max
Load reg. at 10-90% ΔU
OUT
* ΔI
OUT
< 0.75% P
Max
< 0.75% P
Max
< 0.75% P
Max