PS 2000 B Triple Series

3.4.9  Series connection

The outputs 1 and 2 can be connected in series. Series connection is done to achieve a higher output voltage, 

which doubles in this case. The wiring is simple and quickly done using three cables:

In the actual series operation it can be decided to adjust the two outputs separately, but it’s more comfortable to 

use tracking mode (see 

«3.4.8. Tracking mode»

). There are no special things to consider for series operation, ex-

cept that when using a 42 V model in series connection, which then can make up to 84 V, the total output voltage 

would be higher than the SELV limit of 60 V.

3.4.9.1 

Restrictions

• 

There is no totals formation of the voltage; both displays must be read and their actual voltage added

• 

When tracking mode isn’t used, the serial current is limited by the output with the lower current set value

• 

Series connection of multiple devices isn’t permitted, because the total voltage could exceed the voltage withstand 

of the internal insulation between DC and ground (PE). 

3.4.10  Parallel connection

The output 1 and 2 can be connected in parallel. Parallel connection of multiple devices is also possible, but limited 

by the max. current rating of those laboratory cables with 4 mm plugs unless custom cable and 20 A are used. 

The current capability of the DC output sockets (max. 25 A) is also something to consider. We recommend to only 

put models with a lower current rating in parallel, like the one with 3 A where two units could already achieve 12 

A. With the 5 A model it’s already 20 A with two units.

Never connect output 3 in parallel to any other output, because these can generate much higher 

voltage and could damage output 3!

Parallel connection is used to increase the total output current, which in this case doubles. However, the device 

wouldn’t do totals formation on any of the displays. The wiring is simple and quickly done using four cables:

Содержание 39 200 120

Страница 1: ...Elektro Automatik PS 2000 B TFT Triple DC Power Supply Operating Manual Doc ID PS2TEN Revision 14 Date 11 2020...

Страница 2: ......

Страница 3: ...res before installation and use 14 2 3 2 Preparation 14 2 3 3 Installing the device 14 2 3 4 Connection to DC loads 15 2 3 5 Grounding of the DC outputs 15 2 3 6 Connecting the USB port 15 3 OPERATION...

Страница 4: ...as well as general notices in this document are shown in a box with a symbol as follows Symbol for a life threatening danger Symbol for general safety notices instructions and damage protection bans S...

Страница 5: ...B Construction Version B Second generation Maximum current of the device in Ampere Maximum voltage of the device in Volt Special marking 3 Three outputs triple Series 2 Series 2000 Type identification...

Страница 6: ...ulations must be applied when plugging interface cards or modules into the relative slot Interface cards or modules may only be attached or removed after the device is switched off It s not necessary...

Страница 7: ...he responsibilities for operation maintenance and cleaning of the equipment ensure that all employees who use the equipment have read and understood the manual Furthermore the users are to be regularl...

Страница 8: ...pair of DC output sockets and is called output 2 The upper pair of DC output sockets form output 3 The meaning of the signals on the display of output 1 and 2 is as follows Signal OT OverTemperature...

Страница 9: ...P 24 mVRMS Output 1 2 Current Adjustment range 0 6 A 0 10 A 0 3 A 0 5 A Accuracy 1 at 23 5 C 0 3 IMax 0 3 IMax 0 3 IMax 0 3 IMax Load regulation at 0 100 UOUT 0 15 IMax 0 15 IMax 0 15 IMax 0 15 IMax O...

Страница 10: ...Automatik GmbH Helmholtzstr 31 37 41747 Viersen Germany Fon 49 2162 3785 0 Fax 49 2162 16230 www elektroautomatik de ea1974 elektroautomatik de PS 2000 B Triple Series 1 8 4 Views Figure 1 Front side...

Страница 11: ...alue or adjustment of parameter value in the menu Push selection of the decimal position cursor of the currently assigned value 7 Pushbutton On Off Used to switch the DC output oon or off unless the H...

Страница 12: ...es For these devices the following accessories are available Multi Control license Ordering numbers Single license 33 100 229 5 pack of licenses 33 100 230 The device is delivered with an USB stick th...

Страница 13: ...tatus display lower part This area displays various status texts and symbols Display Description The HMI is locked The HMI is unlocked Remote The device is under remote control via USB Alarm xxx Alarm...

Страница 14: ...3 1 Safety procedures before installation and use Before connecting to the mains ensure that the connection is as shown on the product label Overvoltage on the AC supply can cause equipment damage In...

Страница 15: ...ding of the DC outputs The metal socket in proximity to output 1 is internally wired to PE and can be used to ground one of the DC output poles of all three outputs by direct connection or it can be u...

Страница 16: ...e Voltage regulation is also called constant voltage operation CV The DC output voltage of a power supply is held constant on the adjusted value unless the output current reaches the adjusted current...

Страница 17: ...will be in constant voltage mode If however the actual current reaches the set value of current the device will switch automatically to current limiting While the DC output is switched on and constan...

Страница 18: ...ltage alarm OVP will switch off the DC output and can occur if the power supply itself as a voltage source generates an output voltage higher than set for the overvoltage alarm threshold OVP 0 110 UNo...

Страница 19: ...or a blackout doesn t leave the device enough time to store the latest values It s practically powered down immediately Thus the device stores the set values and all settings automatically when they...

Страница 20: ...left HMI Element Description In right HMI Language Selection of the display language between German and English Default setting English No Backlight The choice here is whether the backlight remains Al...

Страница 21: ...bs 1 With the main display being shown it means no menu mode is active turn the left hand knob to adjust the output voltage and the right hand knob to adjust the output current This doesn t depend on...

Страница 22: ...How to lock the HMI with PIN Don t activate the PIN lock if you are unsure about the current PIN It can be changed but only if the current PIN is entered Resetting the device via remote command will a...

Страница 23: ...ivated tracking is saved along with other device settings and thus instantly restored after powering the device Even in tracking mode the two outputs can have separate alarms which would always lead t...

Страница 24: ...t used the serial current is limited by the output with the lower current set value Series connection of multiple devices isn t permitted because the total voltage could exceed the voltage withstand o...

Страница 25: ...considered completely differently to the other outputs It s meant to be an auxiliary output with a fixed voltage which is barely adjustable ca 3 to ca 6 V with a trimmer That trimmer is placed between...

Страница 26: ...can try to acknowledge and delete the alarm by pressing button On Off In order to acknowledge an alarm during remote control see the programming guide These device alarms are configurable Alarm Meani...

Страница 27: ...detail as possible is attached if shipping destination is abroad the necessary customs papers are attached 4 2 1 Replacing a defect mains fuse The device is protected by a fusible which is inside a f...

Страница 28: ...EA Elektro Automatik GmbH Co KG Development Production Sales Helmholtzstra e 31 37 41747 Viersen Germany Fon 49 2162 37 85 0 Telefax 49 2162 16 230 ea1974 elektroautomatik de www elektroautomatik de...

Отзывы: