54
© 2006, Elektro-Automatik GmbH & Co. KG
Irrtümer und Änderungen vorbehalten
EN
Instruction Manual
PSI 8000 3U Series
Date: 05-03-2012
About the power supply
The state of the power output is displayed in the bottom right
corner of the display.
{ON,OFF}
State of the power output
The presently active control mode is displayed to the right of
the related actual values. For instance, the abbreviation „CV“ is
displayed next to the actual value of voltage, because it means
that „Constant voltage“ mode is active. The output values are
limited by the active control mode:
- limited by the voltage set value
(= Constant Voltage)
- limited by the power set values
(= Constant Power)
- limited by the set value of current
(= Constant Current)
- limited by the set value for internal resistance
(optional at U/I/R mode), indicated next to the
actual voltage
(= Constant Resistance)
Additionally to the state of the output an alarm, a warning or a
signal can be displayed:
Alarm
Example:
= Overtemperature
Warnings
Example:
= Overvoltage
Signals
Example:
= Overcurrent
The location from where the unit is currently controlled is dis-
played below the output state. This location is absolute, which
means that you cannot control the unit from elsewhere without
changing the location.
local
Control only possible at the unit
remote
Remote control via communication interfaces
(IF-C1, IF-R1, IF-U1 etc.)
extern
Remote control via built-in or optional,
analogue interface
6.4 Switching the power output on
By pressing the
ON
button the power supply output
is switched on, as long as it is not overridden by the input pin
„REM-SB“ (13) of the built-in analogue interface or optional ana-
logue interface card IF-A1, because the pin has higher priority.
If so and when trying to switch the output on by the button, the
display will indicate the status text „
auto ON
“, noticing the user
that the output will switch on as soon as the override from the
pin is removed.
The output state is indicated in the display with
ON
.
Note: in
local
state (see section 6.9), the pin REM-SB of the
analogue interface (internal or external) is inoperative.
The display shows the current state with „
ON
“.
The
OFF
button switches the power supply output
off. This state is displayed with
OFF
.
The display shows the current state with „
ON
“.
The
OFF
button switches the power supply output
off. This state is displayed with „
OFF
“.
6.5 Adjusting set values
As long as „
extern
“ or „
remote
“ are not shown in the display,
the set values for voltage, current or power can be set manually.
The mode is selected in the device setup at
Accept set va-
lue.
The setting can be accessed with
->
Profile
->
General settings
->
Control panel
the control panel“ for details.
Direct setting of the set values
Using the rotary knobs directly sets the set values.
The left rotary knob adjusts the voltage. The set value of the
voltage is displayed invertedly while it is selected and adjusted.
The right rotary knob either sets the set value for the current, for
the power or internal resistance (optional, unlockable, with U/I/R
mode chosen). The selected set value is displayed invertedly.
With the
SELECT
keys
the set value for the power, with
the set value for the internal resistance or with
the set value for the current is selected.
The maximum adjustable power can also be limited.
Set values are submitted
Alternatively to the direct adjustment of set values you can
choose to set the set values only after submitting them with
the
RETURN
button. See section „7.4 Configuring the control
panel“ for details. The set values can still be changed with the
rotary knobs, but are not set to the output as long as they’re
not submitted. While the set value is unchanged, only its unit
is displayed invertedly. If the set value is changed it is also
displayed invertedly.
The
SELECT
keys switch from current adjustment to power
adjustment for the right rotary knob. The chosen set values are
not submitted to and set by the power supply until then.
Pressing the
RETURN
button submits the set values.
Pressing the
ESC
button discards the new set values
and the old set values are displayed again.
Note: the adjustment of the resistance set value is only
accessible after the optional „internal resistance control“
is unlocked (see section 7.8).
The resistance set value is adjustable from 0Ω up to
20* Unom/Inom. Means, for example, at a device with Unom
= 80V and Inom = 510A it can be adjusted to a maximum
of 3.13Ω.
Содержание 09 230 430
Страница 2: ......
Страница 13: ...13 PSI 8000 3U 系列产品说明书 CN 日期 05 03 2012 关于产品 图 5 ...
Страница 38: ......
Страница 49: ...49 Instruction Manual PSI 8000 3U Series EN Date 05 03 2012 About the device Figure 5 ...