70
© 2006, Elektro-Automatik GmbH & Co. KG
Irrtümer und Änderungen vorbehalten
EN
Instruction Manual
PSI 8000 3U Series
Date: 05-03-2012
Operating the device
Attention! Never connect grounds of the analogue
interface to minus (negative) output of an external
control application (PLC, for example), if that control
application is otherwise connected to the negative
power supply output (ground loop). Load current may
flow over the control leads and damage the device!
Figure 11
Figure 12
10.3 Settings in the device setup
+
Analogue int
The setup menu gives access to some settings for the built-in
analogue interface:
Analogue voltage
Default:
0...10V
= { 0...5V | 0...10V }
Selects the voltage range for the
analogue in- and outputs, which cor-
responds to 0...100% set/actual value
REMOTE /5
Default:
LOW
REM-SB /13
Default:
LOW
= { LOW | HIGH}
Defines, if the digital inputs will react
their dedicated function with either LOW
or HIGH level. For the levels of LOW
Note: both pins are internally pulled-up to HIGH. It means,
that if setting HIGH is selected and nothing is connected to
the analogue interface, pin REMOTE would instantly switch to
permanent remote control and pin REM-SB would switch the
output permanently off.
OVP /14
Default:
LOW
OT /6
Default:
LOW
CV /6
Default:
LOW
= { LOW | HIGH}
Defines, if the digital outputs will report
their dedicated status with either LOW
or HIGH level. For the levels of LOW