6-4 Measuring Temperature
Temperature Measurement for outdoor:
• Set the function switch to the green "0.1°C" position or "1°C" position.
•Then the display will show the environment temperature reading value °C
directly.
• Insert the black plug of temperature probe the COM jack and red plug to the
"V/Hz%/
Ω
/CAP/°C" jack.
•Touch the end of the temperature sensor to the area or surface of the object
to be measured. The display will show the temperature reading value °C
directly.
6-5 Measuring DC Voltage
•Insert the black test lead banana to the COM jack and red test lead banana to
the "V/Hz%/
Ω
/CAP/°C" jack.
•Set the function switch to the green at DCV ranges to be used and connect
test leads connect test leads across the source or load under measurement.
•Set the function switch at DCmV ranges to be used connect test leads
connect test leads across the source or load under measurement.
•Read LCD display. The polarity of red connection will be indicated when
making a DC measurement.
• Press the Hz% button to indicate "Hz".
• Read the frequency in the display. s Press the Hz% button again to indicate
"%".
• Read the % of duty cycle in the display.
6-6 Measuring AC Voltage
• Insert the black test lead banana to the COM jack and red test lead banana
to the "V/Hz%/0/
Ω
/CAP/°C" jack.
•Set the function switch to the green at AC ranges to be used and connect test
leads connect test leads across the source or load under measurement.
•Read LCD display. The polarity of red connection will be indicated when
making a AC measurement.
•Press the Hz% button to indicate "Hz".
•Read the frequency in the display. Press the Hz% button again to indicate
"%.
•Read the % of duty cycle in the display.
6-7 Measuring DC Current
• Insert the black test lead banana plug into the negative COM jack, and the
red test lead banana plug into the "uA/mA"or "10A" jack.
• For current measurements up to 4000uA DC, set the function switch to the
µA position and insert the red test lead banana plug into the uA/mA jack.
•For current measurements up to 400mA DC, set the function switch to the
mA position and insert the red test lead banana plug into the µA/mA jack.
•For current measurements up to 10A DC, set the function switch to the yellow
10A position and insert the red test lead banana plug into the 10A jack.
•Press the MODE button to indicate "DC" on the display.
•Remove power from the circuit under test, then open up the circuit at the
point where you wish to measure current.
•Touch the black test probe tip to the negative side of the circuit. Touch the
red test probe tip to the positive side of the circuit.
•Apply power to the circuit.
•Read the current in the display.
6-8 Measuring AC Current
•Insert the black test lead banana plug into the negative COM jack. and the
red test lead banana plug into the "uA/mA"or "10A" jack.
•For current measurements up to 4000uA AC, set the function switch to the uA
position and insert the red test lead banana plug into the uA/mA jack.
•For current measurements up to 400mA AC, set the function switch to the mA
position and insert the red test lead banana plug into the uA/mA jack
•For current measurements up to 10A AC, set the function switch to the yellow
OA position and insert the red test lead banana plug into the 10A jack
•Press the MODE button to indicate "AC" on the display.
•Remove power from the circuit under test, then open up the circuit at the
point where you wish to measure current.
•Touch the black test probe tip to the neutral side of the circuit.
•Touch the red test probe tip to the "hot" side of the circuit.
•Apply power to the circuit. -Read the current in the display.
•Press and hold the Hz% button to indicate "Hz".
•Read the frequency in the display.
•Momentarily press the Hz% button again to indicate "%".
•Read the % duty cycle in the display.
•Press and hold the Hz% button to return to current measurement.
6-9 Measuring Capacitance
•Set the function switch to the green CAP position.
•Insert the black test lead banana plug into the negative COM jack and the red
test lead banana plug into the "V/Hz%/
Ω
/CAP/°C" jack. (If value is no zero in
the display. Press the REL button to zero)
ELCART DISTRIBUTION SPA via Michelangelo Buonarroti, 46 - 20093 Cologno Monzese (Milano) ITALY