30
3. MAKEREADY
3.12.1
SWITCH ON HV/IS
The diagram shows the states for the HV-test. All prerequisites must be
met for the high voltage test (see also state table further forward in this
document). The safety circuit(s) must be closed.
IN THE „HV ON“ STATE THE HIGH VOLTAGE IS APPLIED.
THERE IS MORTAL DANGER WHEN WORKING WITH HIGH VOLT-
AGE! FOR THIS REASON: WORK PARTICULARLY CAREFULLY
AND CAUTIOUSLY!
Note:
If the „SET“ and „RESET“ signal (X7, pin 3) are present at the same
time, the „RESET“ command has priority: there is no start.
See also chapter 4
OPERATING MODES.
Fig. 3.15: 1. and 2 starts mode
Red
HV ON
Ready for
switch on
Ready
2 Start Mode
Tester on
Digital Start X7
No
Tester is waiting for
2. Start
High Voltage ON
2. Start
via RS232
$SST
Chancel by
Reset-Button
or
$SHT
Test time finish
Yes
1. Start
via RS232
$SST
Tester is waiting for
2. Start
2. Start
via RS232
$SST
Set up device and test
parameter
Test object
faulty
Yes
1. Start
via RS232
$SST
Set remote control
$GOREM
Set remote control
$GOREM
Set communication
$SCM1
No
No
yes
No
No
No
No
Yes
Yes
Yes
Yes
1 Start Mode
1. Start
(Set-Button)
No
Tester is waiting for
2. Start
Set Local mode
$GOLOC
No
Yes
2. Start
(Set-Button)
High Voltage OFF
Green
Full-electronic G1-1 HV DC-Tester
Содержание G1-1