91
11. TEST PARAMETER
Full-electronic G1-1 HV DC-Tester
Start voltage
Test voltage
U
(V)
t (s)
Run back (on/off)
- activates the ramp-down function of the test voltage.
Off
- no voltage ramp down. The full test voltage is switched off
abruptly. In case of test objects with integrated inductance, this
action can result in switching off voltage spikes.
On
- voltage ramp down active. The test voltage is reduced lin-
early from the test voltage to the start voltage as per the ramp
down speed and then switched off.
Ramp down speed (0.01 ... 1.00 kV/s)
- the test voltage is re-
duced linearly by the value given per time unit.
Start up voltage (on/off)
- defines the voltage value from which the high
voltage ramp starts.
on (0.04 kV/s up to (test voltage - 0.01 kV/s))
off
- start voltage - No ramp function, the full test voltage is
switched on.
Test time (Continuous / 0.5 ... 999.9 s)
- the test time can be set from 0.5
to 999.9s. The test time is irrelevant in the case of „Continuous test on“.
Continuous test
-active. The test voltage is applied continuously.
The test must be interrupted manually using the red „Reset“
button S2/H2 if a fault occurs (eg. I > Imax). The continuous
test can last max. 999.9 seconds.
Test time
0.5 ... 999.9 s
- test time active. The test voltage is ap-
plied for the duration of the test time.
Fig. 11.1b: Test voltage ramp