13 Using the valve
The shut-off valve serves to make or break the connection between the cell volume and the
dead volume of the dilatometer when using a specialized frit flange. When running the
experiment, the valve should be open. This way, unwanted pressure build-up via gas evolution
is effectively mitigated.
NOTES
:
•
Close the valve gently by hand. Excessive torque may damage the valve.
•
Some valve parts may get into electrolyte contact. It is therefore advised to unscrew
the valve seating and the valve stem after use. The two O-ring seals and the PTFE
ferrule may stay in place. Wash with plenty of water or other appropriate solvent.
•
Dry the valve parts in the disassembled state (80°C, vacuum, overnight) before
reassembly inside the glove box.
Components of the valve:
Ferrule 1.0
ECC1-00-0029-B
O-Ring 5x1 mm (AP370)
DIC9036
O-Ring 2x1 mm (AP312)
DIC9037
Valve Seating ECD-3
ECC1-00-0153-C
Valve Stem ECD
ECC1-00-0063-B